Frequently asked questions 

 

MAM 2in1 Single Breast Pump  

This chapter lists the questions most frequently asked about the device.  
 

 

Problem 

Solution (Answers) 

1.

 

What do I do if 
there is low or no 
suction when 
pumping with the 
MAM 2in1 Single 
Breast pump? 

 

Check that all connections are properly attached at all 
connection points. 

 

Check that the silicone cushion is firmly seated in the funnel 
adapter. 

 

Check that the valve body is firmly connected to the funnel 
adapter. 

 

Check that the valve membrane is securely seated and flat 
on the valve body. 

 

Check that the membrane cap/handle holder is properly 
connected. 

 

Check that the tubing is not bent and there is no liquid in 
the tubing. 

 

Ensure that the breast shield forms a complete seal around 
the breast. 

2.

 

Milk backflow  

 

If milk or water is found in the membrane or tubing, rinse 
the tubing with clean, hot water. 

 

Hang up to air dry completely before using. 

 

Check to ensure that the membrane is not damaged. 

 

If E1 or E2 appears on the LED display, it means that there is 
an open or short circuit in the internal battery. Please 
contact the MAM Customer Service department. 

3.

 

If the monitor is 
not running

 

 

 

Check to see that the power cord is properly attached to 
the pump if the pump is running on battery. 

 

Check to see if the toggle switch is ON without the adapter. 

 

Ensure battery power is available without the adapter. If 
not, plug in for charging or pumping. 

4.

 

How do I clean the 
tubing  

Inspect the tubing after each pumping session.  
Do not clean the inside of the tubing as no milk should enter it. 
You can clean the tubing in warm soapy water with a damp 
cloth or sponge after and rinse thoroughly with cold water 
afterwards.   

5.

 

What do I do if 
milk is in the tube?   

1.

     

Turn off breast pump 

2.   Unplug breast pump from power source. 
3.   Remove tubing by pulling straight out of tubing ports. Do not 

wiggle or pull tubing at an angle. 

4.   Remove tubing from membrane cap. 
5.   Rinse tubing in cool water to remove breast milk. 
6.   Wash tubing in warm soapy water. 
a.   Rinse tubing with clear water. 
b.   Hang to air dry. 

6.

 

Does the 2in1 
Single Breast Pump 

The 2in1 Single Breast Pump has a closed system since the 
pumping membrane is the barrier.  