MAINTENANCE INTERVALS
Every two years and after every emergency lowering
§
Replacement of the two 9 V block batteries
§
If the 9 V block batteries are not replaced during the
maintenance interval or after use, the electric emergen-
cy lowering function will cease to work.
§
The hospital bed can still be used and fully complies
with the applicable standards and guidelines.
Annually
§
Inspection and maintenance
As required
§
Lubrication of mechanical components
§
Replacement of worn components