Drives cannot
be operated
using the hand
controller
Mains cable is not
plugged in
Plug in the
mains cable
Socket not live
Check socket
Cable plug con-
nection not firmly
connected
Check plug con-
nections on the
motor and hand
controller
Hand controller or
drive is faulty
Inform the op-
erator, specialist
retailer or Betten
Malsch GmbH
Mains cut-off not
activated
(only for
ICS control units)
Activate the mains
cut-off by pressing
the green button
Functions locked
on the hand
controller
Enable functions
on the hand
controller
Motorised ad-
justment system
is not functioning
properly
There is an
obstruction in the
adjustment area
Check moving
parts and remove
any obstructions
The safe working
load has been
exceeded
Reduce the load
Drives cut out af-
ter a long period
of operation
The adjustment
time or safe
working load has
been exceeded
and the control
unit has reacted to
overheating
Allow the drive
system to cool
down sufficiently.
Opposite function
activates when
operating the
hand controller
buttons
Motor plugs have
been mixed up
Check that the
cables are con-
nected correctly
or inform your op-
erator, specialist
retailer or Betten
Malsch GmbH
The side rails
can no longer be
properly adjusted
The mechanism is
obstructed or bent
Check all moving
parts and remove
any obstructions,
or contact our
customer service
team
Castors do not
brake or cannot
be rolled
Obstructions have
become trapped
in the castors
Remove obstruc-
tions
The castor system
is faulty
Contact our
customer service
team
T R O U B L E S H O O T I N G