Ground balance can be performed in three ways with the
Kruzer
: Automatic, Manual and
Tracking.
When the GB button is pushed while performing automatic or manual ground balance,
the device will switch to the General Search (GEN) mode automatically on the background
without any indication to the user, regardless of the selected search mode.
Upon completion of ground balance, current ground balance value is shown in the Ground
Balance (GB) box on the right side of the display.
Automatic Ground Balance
Automatic ground balance is performed as follows in all search modes:
1)
Find a spot where there is no metal.
2)
Push and hold the GB button down (GROUND BALANCE value and “Pump Coil” warning
message will be shown on display) and start pumping the search coil up and down from
about 15-20 cm (~6''- 8'') above the ground down to 3 cm (~1'') off the ground with smooth
movements and keeping it parallel to the ground.
3)
Continue until a beep, indicating the completion of ground balance, is heard. Based on
ground conditions, it usually takes about 2-4 pumps for the ground balance to be completed.
4)
Upon completion of the ground balance, ground balance value is shown on the display
(GB). The device continues to ground balance and produce a beep sound as long as you keep
the GB button pushed down and pump the coil. In order to ensure that the ground balance is
proper, ground balance at least 2-3 times and check the ground balance values on the display.
In general, the difference between the values shall not be higher than 1-2 numbers.
5)
If you cannot ground balance, in other words, if no beep sound is produced, it means that
either the ground is too conductive or not mineralized or there is a target right below the
search coil. In such a case, retry ground balance at a different spot. If you still cannot ground
balance, read the section titled
''Important Details Concerning Ground Balance''
.
When the ground balance button is released, the device continues to operate in the GEN
mode for a short period of time and the ground balance value stays on display. This makes
it possible to manually fine tune the automatic ground balance value. Refer to the following
"Manual Ground Balance"
section for further information regarding this feature. If this is
not desired, press the PP button once to return to the main screen.
NOTE:
If the iSAT value is set high, the device may not auto ground balance. In such a case,
lower the iSAT value first. After ground balancing, set the iSAT back to its original position.
Page 19
Manual Ground Balance
Allows you to manually modify the ground balance value. It is not preferred mostly because
it takes time. However, it is the preferred option in cases where a successful ground balance
cannot be performed using other methods or minor corrections are required to the
automatic balance.
Kruzer
is designed to allow for automatic ground balancing conveniently on any type of
ground. Therefore, it is recommended to perform automatic ground balance upon start up.
However, the ground may not be suitable for automatic ground balancing in some cases
and the device cannot ground balance on such grounds (Except for the BEACH mode). For
instance, wet beach sand, soils containing alkali or salty water, trashy sites, ploughed fields,
highly mineralized grounds and grounds with very low mineralization are not suitable for
automatic ground balance. In such terrains, you can auto ground balance in the BEACH
mode and then switch to other modes or try manual ground balancing. However, manual
ground balance requires a skill which develops over time through practice.
To perform manual ground balance:
1)
Find a clear spot without metals and switch the device to the GEN mode.
2)
You need to listen to the sounds coming from the ground in order to perform manual
ground balance. Pump the search coil up and down from about 15-20 cm (~6''- 8'') above
the ground down to 3 cm (~1'') off the ground with smooth movements and keeping it
parallel to the ground.
If the sound gets higher when lifting off the search coil above the ground, the ground
balance value is too low, in other words, the effect from the ground is negative and the
ground balance value needs to be increased by using the ( + ) button. On the other hand,
if the sound gets higher when lowering the search coil to the ground, the ground balance
value is too high, in other words, the effect from the ground is positive and the ground
balance value needs to decreased by using the ( - ) button.
3)
Push the ground balance button once and release it. The ground balance value will be
shown on the display and remain there for a moment. You can return to the ground balance
screen by pushing the ground balance button if the screen switches.
Manual ground balance functions within the
range of 0-99.80 However, each value covers
5 steps used for fine tuning within itself and
these steps are indicated as multiples of 20 in
the Ground Balance window (GB). For example,
ground balance value shown on the side is 70.80.
Press ( + ) or ( - ) to increase or decrease the ground
balance value, respectively. If the key is pressed
once at a time, the values count one by one and if
it is held down, the values will change quickly.
4)
Repeat the above procedure until the sound heard from the ground is eliminated.
The sound may not be eliminated completely in some areas. In these cases, listen to the
sounds produced when moving the search coil towards and away from the ground to check
if the ground balance is correct. If there is no difference between the two sounds then the
ground balance is set properly.
The device will return to the main screen automatically after a short period of time upon
completion of ground balance. To return to the main screen instantly, just press the PP
button once.
IMPORTANT!
Experienced detectorists adjust the ground balance setting to a slightly posi-
tive response (weak but audible sound is produced when moving the search coil closer to
ground). This method may produce favorable results for experienced users in certain fields
where small targets are searched for.
Ground Tracking (TRACKING)
In this option, the user does not need to make any adjustments. TRACKING feature is
activated from the OPTIONS menu by switching it to 01 position. The word ''Tracking''
is displayed at the bottom of the GB window. The device updates the ground balance
automatically as long as the search coil is swung over the ground and shows the ground
balance value in the GB window. It does not provide any feedback to the user (like the
beep sound in automatic ground balance).
While tracking is active, the device can initially produce a loud signal when it detects a
different ground structure (for instance a mineral rock) or a target. In this case, swing the
search coil over the spot where the device produces the signal . If the sound remains the
same and the device shows an ID, it is possibly a target. If the sound attenuates too much
or is lost after a few swings, it means that the device has produced a signal for the different
ground structure or a stone.
NOTE:
It is recommended that you use tracking in the General Search mode (GEN) and not
in the discrimination modes.
Tracking is suitable for use in areas where different soil structures are present within the
same land or in fields where mineralized rocks are scattered widely apart. If you use ground
tracking in areas where hot rocks are intensely present, the device may not be able to elimi-
nate these highly mineralized rocks or you may miss the smaller or deeper metals.
IMPORTANT!
Ensure that tracking is off during air tests. Otherwise, the device will attempt
to perform ground balance on the target and the depth will be reduced.
Ground Balance Value
Ground balance value provides information about the ground you are searching on. Some
typical ground types are as follows:
0-25
Wet salt water or wet alkali soils
25-50 Wet salt water and wet alkali soils covered with dry layers
50-70 Regular, low-quality soils
70-90 Highly magnetic soils, magnetite or maghemite and similar highly mineralized soils,
black sand.
Important Details Concerning Ground Balance
1)
Upon start up, the ground balance value is set to 90. The device can perform ground
balance automatically within the range of 20-99.80 in all modes and 00-99.80 in the BEACH
mode.
2)
If the ground mineralization is too low, automatic ground balance may fail to work in
other modes except for the BEACH mode. In such a case, you can auto ground balance in
the BEACH mode and then switch to other modes or try manual ground balancing.
3)
You can test the accuracy of the ground balance with the pinpoint mode. After ground
balancing, if you receive no sound or a weak one when you move the search coil closer to
the ground in the pinpoint mode, then the ground balance is successful. If the sound gets
louder when you move the search coil closer to the ground, then the ground balance is not
successful. In this case, simply change your location. If ground balance is not possible despite
these efforts, you should continue your search without performing ground balance.
You cannot search in the General Search (GEN) mode without ground balancing. You need to
use one of the discrimination modes and increase the DISC. value until the noise is eliminated.
4)
Once the ground balance is set, it will remain satisfactory for a long time in most areas.
However, if you encounter an excavated, backfilled or geologically composite soil structure,
a ground balance should be performed again to adapt to the varying soil structure. In
addition, re-ground balancing is recommended for the multi Kruzer model if you change
the operating frequency of the device (5kHz/14kHz/19kHz) in certain ground conditions.
5)
When using the optional large coil, pump the coil more slowly and do not keep it very
close to the ground.
6)
In some cases where the iSAT value is set high, the device may not be able to ground
balance automatically. In such a case, first lower the iSAT and after ground balancing switch
it back to its previous position.
TARGET ID is the number produced by the metal detector based on the conductivity of the
metals and gives an idea to the user about what the target may be. Target ID is shown with
two digits on the display and ranges between 00-99.
NOTE:
Keep in mind, large targets will ID higher than expected, even though they may be
of lower conductance.
In some cases, the device may produce multiple IDs for the same target. In other words, the
IDs may be jumpy. This may result from several factors. Target orientation, depth, purity of
the metal, corrosion, mineralization level of the soil etc. Even the direction of the search coil
swing may cause the device to generate multiple IDs.
In some cases, the device may fail to provide any ID. The device needs to receive a strong
and a clear signal from the target in order to provide an ID. Therefore, it may not be able to
provide an ID for targets at fringe depths or smaller targets even if the device detects them.
Keep in mind that target IDs are “probable”, in other words, estimated values and it would
not be possible to know the properties of a buried object exactly until it is dug out.
IDs of non-ferrous metals such as copper, silver, aluminum and lead are high. Target ID range
of gold is wide and may fall within the same range of metal wastes such as iron, foil, screw
caps, and pull tabs. Therefore, if you are looking for gold targets, digging out some trash
metals is expected.
Push the SETTINGS button to access the basic settings. You can navigate the basic settings
with the up and down buttons. The value of the selected setting will be displayed on
screen. You can change the value using the plus (+) and minus (-) buttons. If the up/down
and +/- buttons are held down, the options and values will change rapidly.
To exit the settings, press the SETTINGS or the PP button once. Settings will time out in
approximately 8 seconds and the device will revert back to the modes window.
NOTE:
Certain settings are mode specific and thus cannot be selected in other modes.
GAIN
Gain is the depth setting of the device. It is also used to eliminate the ambient electromag-
netic signals from the surrounding environment and noise signals transmitted from ground.
NOTE:
To obtain maximum depth performance, to eliminate the noise caused by electro-
magnetic interference, try shifting the frequency first.
Frequency shift is done with the FREQ. option in the
Kruzer
and with a combination of
buttons in the
multi Kruzer
(see pages 21-22)
. In the
multi
Kruzer
, when the frequency
shifting is not sufficient for eliminating noise, you can also change the operating frequency
(5kHz/14kHz/19kHz) of the device.
Gain setting range is 01-99 and pre-defined for each mode. All modes start at default
settings. They can be manually modified when necessary. Gain adjustment applies to the
selected mode; the modified setting does not affect the gain setting of the other modes.
NOTE:
If the ground is highly mineralized causing the device to overload, decrease the gain
until the ''Overload'' message disappears from the screen.
Gain in General Search Mode (GEN):
In the GEN mode, gain setting causes an increase or decrease in the popping sounds and false
signals. Gain setting is a personal preference. However, It is important to set the gain setting
to the highest level possible where no major popping sounds are heard to avoid missing
smaller and deeper targets. For example; if the noise level is suitable for searching and is the
same at gain levels 40 and 70, then 70 should be preferred. Using the factory default levels
will be a good starting point until you get familiar and experienced with the device.
Gain in Discrimination Modes:
Since the threshold setting is not available in the discrimination modes, you can increase
the depth of the device or ensure noise-free operation on different grounds only by using
the GAIN setting.
In order to adjust the gain in the discrimination modes, first ground balance while the gain
is at its default setting. After ground balance is completed, hold the search coil stationary or
swing over the ground at search height. Reduce the gain if the device receives noise. If not
(ensure that the DISC. is also at its default settings when checking this), increase the gain
gradually until there is no popping sound. If the device starts to receive noise during
searching, reduce the gain gradually.
NOTE:
Kruzer
is a high gain device and some of the search modes will run relatively noisy
(DEEP and 4 TONE) compared to other modes so as to provide the best depth performance.
However, due to the design characteristics of these modes, the noise will be heard more if
the coil is in free air than sweeping the coil on the ground. Please keep this factor in mind
while adjusting the gain.
Discrimination (DISC.)
Disc. is the ability of the device to ignore all metals below a certain Target ID. In the DISC.
process, the filtered ID range is shown with lines on the ID scale and every 2 consecutive IDs
are represented with 1 line. For example, if you set the Disc. to 30, 15 lines will be shown
between the 0-30 ID range on the scale and the device will not produce an audio response
for any metals with IDs between 0-30.
DISC. setting is disabled for GEN mode only. For all other modes, the factory default value
will be displayed on screen at start up.
In order to change the DISC. value, select the DISC. option from the SETTINGS and decrease
or increase the value using the plus (+) or minus (-) buttons. Please remember that certain
targets, other than the ones you want to ignore, may also be missed or their signals may
become weaker when using the DISC. setting.
In the case of receiving multiple IDs for the same target - let's say 35 and 55 - due to the orien-
tation of the target or the composition of the metal itself, if you set the DISC. to 40, because
35 will fall in the filtered range, the signal strength as well as the depth may diminish.
NOTE: DISC. setting works inversely proportional to depth up to level 15 in all modes. In
other words, as the DISC. is increased up to 15, stability will increase but depth will be
reduced and vice versa. Above 15 though, both depth and noise will increase.
NOTCH
NOTCH is the ability of the device to discriminate single or multiple Target IDs by not
emitting an audio response for them or giving a low iron tone (please refer to iron tone in
notch).
Although NOTCH may seem similar to DISC. at first glance, these two settings have different
functions. While the DISC. filters out all IDs between 0 and the set value, the NOTCH filters
IDs individually.
With the NOTCH you can reject a single ID or multiple IDs at the same time. This process
does not affect any IDs below or above the selected IDs. For example, you can filter out IDs
between 31-35 as well as 50 simultaneously.
How To Use The NOTCH Setting
When NOTCH is selected from the SETTINGS, first, the current DISC. value will be displayed
on screen and discriminated ID range will be shown on the ID scale with lines. For example,
if the DISC. is set to 15, when you select NOTCH, number 16 will be displayed on screen
corresponding to 8 lines on the ID scale (every 2 consecutive IDs are represented with 1
line). NOTCH cannot be used within the DISC. range. In other words, if the DISC. is set to 15,
NOTCH can only be applied to IDs 16 or higher. If you want to NOTCH IDs 15 or below, first
you need to change the DISC. value.
NOTCH rejects or accepts IDs with the help of the cursor in the middle of the screen. To
move the cursor on the scale, plus (+) and minus (-) buttons are used.
The cursor blinks
while it is moving on the scale.
When you are on the first ID that you want to reject, press
the SELECT button once. This ID is now rejected and it is shown on the screen with a line. If
you want to reject multiple IDs, continue to press the plus (+) or minus (-) button. If
non-consecutive IDs are to be rejected, push the SELECT button once to have the cursor
blink for navigation on the scale and repeat the process above. The cursor will appear
where you left it the next time you use the NOTCH.
To give an example; let's say you want to reject IDs between 20-25 and the cursor is at 10.
Press the plus (+) button until you reach number 20. Then push the SELECT button once.
Number 20 will be marked with a line. When you reach number 25 using the (+) button
again, IDs between 20-25 will be filtered out and they will be shown on the ID scale with 4
lines (every 2 consecutive IDs are represented with 1 line).
To accept back the filtered IDs, select NOTCH from SETTINGS. The cursor will appear where
you last left it. Using the plus (+) or the minus (-) button, select the ID you want to accept
and push the SELECT button. Then, using the plus (+) or minus (-) button again, unfilter the
IDs back in. 1 line will be erased for every 2 consecutive IDs accepted.
Iron Tone in Notch Setting:
This will enable you to get a low iron tone for your notched out target IDs instead of
silencing them. To use this feature, first select Fe VOL. from the SETTINGS and using the
plus (+) button select the iron tone volume between n1-n5. n5 is the maximum level and
the iron volume will be reduced as you go down but it cannot be silenced completely.
NOTCH adjustment applies to the selected search mode only. The change does not affect
the other modes.
IMPORTANT!
In the
multi Kruzer
, if you are using the Standard ID scale and you change the
operating frequency of the device, you may need to re-adjust the NOTCH values according
to the IDs you will get in the new frequency.
Iron Volume (Fe VOL.)
It adjusts or turns off the volume of the low iron tone. It can be adjusted between 0-5 or
n1-n5.
0-5:
5 is the maximum level. As you lower it, the audio response volume the device
produces for ferrous metals will decrease. At 0 level, the iron audio will be silenced. In
other words, the device will detect ferrous targets, the Target ID will be displayed on the
screen but the device will not produce any audio response.
n1-n5:
This will enable you to get a low iron tone for your notched out target IDs instead of
silencing them. n5 is the maximum level and the iron volume will be reduced as you go
down but it cannot be silenced completely.
Fe VOL. adjustment applies to the selected search mode only. The change does not affect
the other modes.
Tone Break (T.BREAK)
It is used to adjust the break points of the target response tones on the Target ID range.
Default Tone Break points in the Kruzer will vary according to the search mode. By using the
Tone Break feature, for each metal group (Fe, Gold/Non-Fe, Non-Fe) you can change the
point where the low tone changes into the higher tone.
To use the Tone Break feature, first select T.BREAK from SETTINGS . The names of the metal
groups mentioned above will appear at the bottom of the screen. The Tone Break point of
the metal group will be shown on the screen numerically while the cursor at the top will
point to it on the ID scale. In some modes, there are 2 tone break points and in some there
are 3. To select the metal group, just push the SELECT button. Selection will be framed. To
change the value of the break point, plus (+) or minus (-) button is used.
To give an example for the above explanation; let's say you are in the 3 TONE mode and you
want to change the Tone Break points. First, select T.BREAK from SETTINGS. Fe and
Gold/Non-Fe will appear at the bottom of the screen and Fe will be framed. The default
value of 15 will also be displayed on the screen. Using the plus (+) or the minus (-) button
change this number to any value you want. Let's say you increased it to 40. Then, push the
SELECT button once to select the Gold/Non-Fe. Let's say you decreased the default value of
66 to 50. In this case, the device will produce a low iron tone for all metals with IDs equal to
or less than 40, a medium tone for metals with IDs 41-50 and a high tone for metals with IDs
greater than 50 (If you have also adjusted the audio tones (TONE), the selected frequency
will apply to the new ID ranges).
Tone Break adjustment applies to the selected search mode only. The change does not
affect the other modes.
IMPORTANT!
If you are using the Standard ID scale in the multi Kruzer and you change the
operating frequency of the device, you may need to re-adjust the Tone Break points
according to the IDs you will get in the new frequency.
TONE
Allows you to change the target audio response tones and the threshold sound according
to your preference. For each metal group (Fe, Gold/Non-Fe, Non-Fe) the frequency can be
adjusted between 150 Hz (15) and 700 Hz (70).
When TONE is selected from the SETTINGS, names of the metal groups mentioned above
will appear at the bottom of the screen and the selected one will be framed. To select
another group, just press the SELECT button. Then use the plus (+) or the minus (-) button
to change the audio frequency.
NOTE:
If you want to change the frequency of the 4th tone in the 4 TONE mode, while Non-Fe
is selected, press the SELECT button once. The frame will disappear and the audio frequency
of the 4th tone will appear on the screen. use the (+) or the (-) button to change it.
TONE adjustment applies to the selected search mode only. The change does not affect the
other modes.
Threshold (Thresh.)
In the General Search Mode (GEN), search is performed with a continuous humming sound in
the background, also referred to as the threshold sound. The loudness of this hum directly
impacts the detection depth of smaller and deeper targets and it is adjusted by the threshold
(Thresh.) setting. If the threshold is set too high, a weak target signal may not be heard. On
the contrary, if the threshold is too low, you give up the depth advantage this setting offers.
In other words, weak signals of smaller or deeper targets may be missed. It is recommended
for average users to leave this setting at its default value and for experienced users to adjust
to the highest level where they can still hear the weak target signals.
Threshold level is directly related to the Gain and iSAT settings. Please be sure to read the
related sections of the manual carefully.
iSAT (Intelligent Self-Adjusting Threshold)
iSAT in General Search Mode (GEN)
For the General Search Mode (GEN) to perform accurately, a stable threshold sound is neces-
sary. You cannot search in the General Search Mode without ground balancing. Changes
that occur in the soil structure and mineralization levels after ground balancing, may cause
a rise or fall in the background hum and disrupt the threshold's stability which will result in
false signals and thus missing signals of small metals. iSAT adjusts the speed that the device
recovers its threshold hum and eliminates the negative effects of mineralized soils.
Increasing the iSAT in high mineralization will enable a more stable operation by avoiding
false signals. This, however, may cause some loss in depth and it is normal.
NOTE:
In high mineralization, if you receive too many false signals without disruption in the
threshold hum, lower the gain first before increasing the iSAT. If the false signals continue,
set the gain back to its original value and increase the iSAT.
If the mineralization is low, you can decrease the iSAT and sweep the coil more slowly for a
deeper detection.
iSAT consists of 10 levels. The device will start at level 6. It is recommended that iSAT should
be increased in high-mineralization and decreased in low mineralization.
iSAT in Discrimination Modes
It is used to eliminate false signals caused by ground noise or hot rocks when searching in
discrimination modes and the available range is between 00-10. Its factory default value is
set to (1). You can change the value using the plus (+) and minus (-) buttons.
If the device receives a lot of false signals due to highly mineralized soil or hot rocks in the
discrimination modes, first re-ground balance. If the false signals continue, lower the GAIN
and check again. In case the false signals still exist, try increasing the DISC. value. Regardless
of all these, if the false signals still exist, first change the GAIN and DISC. values back to their
previous levels. Then, increase the iSAT level until the false signals are eliminated.
At the maximum level of iSAT, false signals will disappear or will be minimized. However, in
some cases, increasing the iSAT will result in loss of depth for certain metals such as
copper.
NOTE:
When detecting in 19kHz, on wet or highly mineralized ground, in order not to miss
smaller high conductive metals (silver, copper etc.) it is recommended not to increase the
iSAT level too high.
NOTE:
iSAT value ranges between 00-10. The factory default is 01. At ''0'', the iSAT feature
will be inactive. If the ground is not highly mineralized or does not contain many hot rocks,
setting the iSAT to ''0'' is recommended.
On the
multi Kruzer
, when the operating frequency is changed (5kHz/14kHz/19kHz), the
Target ID will change as well. This represents the ''Standard'' ID scaling of the device.
IMPORTANT! At start up, multi Kruzer will utilize the ''Normalized'' ID scale and not the
Standard ID scale.
In other words, the IDs will not change upon frequency change and the
device will generate the 14kHz IDs in each frequency. However, based on ground conditions
IDs may vary for certain metals.
If you prefer to see the different IDs produced by each frequency, you need use the
''Standard'' ID scale. To switch to the standard IDs,
while the device is on
, push the plus
(+) and minus (-) buttons at the same time. Letters ''Sd'' will appear on the screen. If you
wish to revert back to the normalized IDs, repeat the same process and letters ''no'' will
appear on the screen.
The table showing the probable IDs for the Kruzer and the multi Kruzer models is placed
at the back of this manual. You can easily detach the page and carry it with you during
your searches.
Coins searched throughout the world are made of different metals and in different sizes in
different geographical locations and historical eras. Therefore, in order to learn the Target
IDs of the coins in a specific region, it is suggested to perform a test with the samples of
such coins, if possible.
It may take some time and experience to make best use of the Target ID feature in your
search area. Different brands and models of detectors produce different target ID numbers.
The numbers vary even more depending on target depth, ground mineralization, and
adjacent metals. But after some practice, you will quickly become comfortable with the
meanings of the
Kruzer
's Target IDs.
Target ID Depth
This setting is not present in the menu.
Adjusts the depth level that the device displays an ID for a detected target. It consists of 3
levels: Hi (High), In (Intermediate), Lo (Low). Factory default is set to ''In''.
The lower the ID depth level is, the higher the ID accuracy and vice versa. At the high level,
the IDs may become jumpy.
To change the ID depth level, press the PP button and the up button simultaneously. Each
time you press the up button, the ID depth level will change.
SETTINGS