Major Science mini pro Instruction Manual Download Page 16

 

16 

Section 6 

Trouble shooting and Maintenance 

 

Many operating problems may be solved by carefully reading and following the 

instructions in this manual accordingly. Some suggestions for troubleshooting are 

given below. Should these suggestions not resolve the problem, contact our 

SERVICE DEPARTMENT or a distributor in your region for assistance. If 

troubleshooting service is required, please include a full description of the problem. 

 

Problem 

Cause 

Solution 

No Display / lights 

No AC power 

Check if the power supply is unplugged, 

or AC power source problem 

AC power cord is not 

connected

 

Check AC power cord connections at both 

ends. Use the correct cords.

 

The fuse has blown 

Replace the fuse 

Repeated fuse 

broken 

Hardware failure 

Contact Major Science service 

department 

Operation stops 

 

Electrophoresis leads 

are not connected to 

the power supply or to 

the electrophoresis 

unit(s), or there is a 

broken circuit in the 

electrophoresis cell 

Check the connections to the power 

supply and on your electrophoresis cell to 

make sure the connection is intact; check 

condition of wires in electrophoresis unit. 

Close the circuit by reconnecting the 

cables. Press 

START/STOP 

to restart the 

run. 

High resistance due to 

tape left on a pre-cast 

gel, incorrect buffer 

concentration, or 

incorrect buffer 

volumes in the   

electrophoresis cell 

Correct the condition by making sure the 

tape is removed from the pre-cast gel, 

buffers are prepared correctly, and the 

recommended volume of buffer is added 

to the electrophoresis unit. 

 

Error message 

Over current(400mA 

limitation reached) 

Check if the buffer concentration is 

appropriate. Excessive buffer 

concentration may cause over current 

issue. To clear the error message, press 

the 

START/STOP

 button again 

Summary of Contents for mini pro

Page 1: ...1 Mini Pro 300V Power Supply Instruction manual Catalog No MINI 300 www majorsci com service majorsci com Version 02A Revised on 2012 8 14...

Page 2: ...Major Science is liable for all missing or damaged parts accessories within 7 days after customer received this instrument package Please contact Major Science immediately regarding this issue If no r...

Page 3: ...Product Description Feature 10 Section 3 Technical Specification 11 Section 4 Installation Instructions 12 Section 5 Operation Instructions 13 5 1 Control interface 13 5 2 Start the operation 14 Sect...

Page 4: ...y recommended for the user to read the following points carefully before operating this equipment 1 Read and follow the manual instructions carefully 2 Do not alter the equipment Failure to follow the...

Page 5: ...2000 meters 5 4 40 operating temperature 6 Pollution degree 2 7 Mains supply voltage fluctuations up to 10 of the normal voltage Avoiding Electrical Shock Follow the guidelines below to ensure safe op...

Page 6: ...cleaning or decontamination methods other than manufacturer s recommendation users should check with the manufacturer s instruction to see if the proposed method will damage the equipment Equipment O...

Page 7: ...ural wear of externally exposed parts of the product are scratched or damaged under normal usage including casing key pad display and accessories etc or 4 The defect is arising from improper installat...

Page 8: ...8 of dirt or dust or mains supply problem thunderstorm activities or infestation by insects or vermin or exposure to abnormally corrosive conditions or operation with extreme heat or humidity...

Page 9: ...n the market Sufficient and accurate output voltages two terminal pairs and its compacted size can deliver accurate and reliable experimental results from one experiment to another Mini Pro 300V Power...

Page 10: ...e run simultaneously saving both time and valuable bench space With a universal voltage rating the mini 300V is also designed and constructed to the most rigorous safety standards Packages including s...

Page 11: ...Constant Current Control Microprocessor controller Terminal Pairs 2 Pairs Timer 1 999 minutes with alarm continuous Safety Device No load detection shrouded plugs and sockets Operation Temperature 4...

Page 12: ...ions Mini Pro 300V Power Supply is actually a pre installed instrument As long as it is placed on a sturdy and level surface in a safe dry place and further connects with well prepared electrophoresis...

Page 13: ...sing RUN Start LED light indicates your current parameter If you are setting a value for Voltage the Voltage LED will be lit b Operation Mode After you start the experiment the LED light next to the R...

Page 14: ...with appropriate voltage 100V to 240V as indicated on the rating sticker near the AC cord on the back of the unit 3 Power on the unit by pressing the ON OFF switch 4 Connect the DC output jacks from...

Page 15: ...he LED light next to the button will light up 10 Press Key again to stop the unit at any time if necessary 11 When the run is completed Press Key to terminate a timed run and Turn the AC power OFF by...

Page 16: ...ment Operation stops Electrophoresis leads are not connected to the power supply or to the electrophoresis unit s or there is a broken circuit in the electrophoresis cell Check the connections to the...

Page 17: ...i com 3 If the unit must be shipped back for repair contact Major Science or the distributor for a Return Authorization Number and shipping instructions The unit will be repaired and returned to you a...

Page 18: ...ck into the compartment 5 Snap the cover closed Maintenance Mini Pro 300V Power Supply uses all solid state components and should require no maintenance or recalibration under normal use If the unit m...

Page 19: ...ackage MJSS Mini300 Package of 2 x MJ 105 S and Mini300 MJRR Mini300 Package of 2 x MJ 105 R and Mini300 MTT Mini300 Package of 2 x MT 108 and Mini300 MJSR Mini300 Package of MJ 105 S MJ 105 R and Min...

Page 20: ...ty is limited to the receipt of reasonable proof by the customer that the defect is embraced within the terms of the warranty All claims made under this warranty must be presented to Major Science wit...

Reviews: