background image

Maiman Electronics

                                                                                                                                

MBL1500B

 

21 

 

 

  SCPI control description 

17.

Enter SCPI mode by using the command “P0704 0800”. 
Exit SCPI mode by using the “SCPI OFF” command. 
As  in  the  standard  protocol,  all  commands  must  be  terminated  with  a  carriage  return  symbol  CR 

(0Dh). 

The  separation  of  command  levels  occurs  with  the  “:”  symbol.  To  request  a  value,  use  the  “?” 

symbol without a space immediately after the parameter. To set the value, the new value is specified with 
a space after the parameter. The separator of the integer and fractional part is the point “.”. In the angle 
brackets  <  >  specify  the  parameters  for  which  you  want  to  specify  a  numeric  value.  In  curly  brackets  {} 
through the separator | specify the possible set of commands.  

For example, line “freq:value {<frequency>|?}” means that the user can either set a new frequency 

value, or request the current. 

Frequency request: “frequency:value?”. 
Setting a new frequency value: “frequency:value 30.5”. 
Some commands have a full and shortened version. The full version is shown in the table, and the 

shortened version is shown by UPPERCASE. For example, frequency value can be requested in two ways: 
“frequency:value?” and “freq:value?”. 

The  device  responds  using  the  command  version  that  was  used  by  the  user,  and  the  parameter 

value is specified in the full version. 

 

Table 17. Available list of SCPI protocol commands 

Commands level 

Description 

SCPI 

OFF 

 

 

Exit SCPI mode

 

FREQuency 

:MIN 

 

Minimum 

Frequency (0,1 
Hz) 

:MAX 

 

Maximum 

:VALUE 

{<0.0>|?} 

 

Value 

DURation 

:MIN 

 

Minimum 

Duration (0,1 
ms) 

:MAX 

 

Maximum 

:VALUE 

{<0.0>|?} 

 

Value 

CURRent 

:MIN 

 

Minimum 

Current (0,1 A) 

:MAX 

 

Maximum 

:VALUE 

{<0.0>|?} 

 

Value 

:REAL 

 

Measured value 

:CALibrate 

{<>|?} 

 

Current set calibration (0,01%) 

VOLTage 

:REAL 

 

Measured value 

Voltage (0,1 V) 

DEVice 

:START 

{?|ON|OFF} 

 

Start/stop 

Device state 

:CMODe 

{?|INTernal|EXTernal} 

 

Internal/external 
current set

 

:SYNC 

{?|INTernal|EXTernal} 

 

Internal/external 
Enable

 

:BLOck 

{?|USE|IGNore} 

 

Allow/deny Interlock

 

:TEMPBlock 

{?|USE|IGNore} 

 

Allow/deny NTC 
Interlock

 

Summary of Contents for MBL1500B

Page 1: ...User Manual Before powering on your driver read this manual thoroughly If you have any doubt or suggestion please do not hesitate to contact us Maiman Electronics Saint Petersburg Russia e mail info m...

Page 2: ...v0 0 1 2020...

Page 3: ...ons and weight 3 6 Electrical characteristics 3 7 Electrical characteristics LD power supply channel 3 8 Electrical characteristics TEC 4 9 Typical Performance Characteristics 4 10 Protections 4 11 Co...

Page 4: ...hannel provides current up to 1 5A with noises no more than 10 15 A The TEC channel provides a temperature change from 12 to 40 C with stability 0 01 C TEC contains integrated self adjusted PID contro...

Page 5: ...nect an external emergency switch Allows to connect an external auto protect device e g over temperature switches The laser can only be turned on with shorted Interlock Soft start The soft start funct...

Page 6: ...emperature is exceeded the LD power supply will be automatically turned off After the temperature returns to normal the LD power supply can be switched on again State after switching on After turning...

Page 7: ...Maiman Electronics MBL1500B 6 Controls 12 Front panel Fig 3 Front panel description Back panel Fig 4 Back panel description...

Page 8: ...e 14 TEC cathode 7 n c 15 TEC cathode 8 n c shield chassis ground How to get started 14 When you turn on the MBL1500 without the connected LD the operation and information on the screen will be incorr...

Page 9: ...etting parameters and settings in detail 15 1 The Main screen Fig 7 The Main screen MBL1500 has a duplicate control using buttons and encoder and or touchscreen Button control will be described at fir...

Page 10: ...4 Control with buttons and encoder Use the Left cursor and Right cursor buttons to select the integer or fractional part of the LD current that you plan to change Turn the encoder to change the curre...

Page 11: ...ou need to set a new value clear the input window using the Clear button If you only want to edit the value use the Back space button to remove unwanted symbols Set the required value using the numeri...

Page 12: ...esponds to the required submenu To return to the QCW Setpoint submenu press the Back button To return to the Main menu press the LD Setpoint button Control with touchscreen To go to the submenu tap th...

Page 13: ...nction button 3 Right cursor Moves the change cursor to the right Function button 4 Control with buttons and encoder or touchscreen The TEC Setpoint submenu has the same control principle as the LD Se...

Page 14: ...the Back button to return to the Main menu 15 8 The Settings LD max current screen Fig 14 The Setting LD max current screen Table 9 The Setting LD max current screen submenus description Button Descr...

Page 15: ...ct the memory cell from Memory1 to Memory9 to save the current values and click the Save button To load values move the selection cursor using the Up and Down buttons to the line with the required mem...

Page 16: ...utton 4 Control with buttons and encoder The Memory Keypad screen is available only with a touchscreen Control with touchscreen To go to the Memory Keypad screen in the Memory screen double tap on the...

Page 17: ...Maiman Electronics MBL1500B 16 15 11 The Info screen Fig 17 The Info screen The Info submenu provides general information Tap on the touchscreen or any button to exit...

Page 18: ...arameter Hex number of the parameter For example 0100h 6 space symbol 20h 7 10 New value of the parameter Hex value of the parameter 0000h 11 return carriage symbol CR 0Dh End of the command The devic...

Page 19: ...0 1 Hz Value R W 0100 Minimum R 0101 Maximum R 0102 Duration 0 1 ms Value R W 0200 Minimum R 0201 Maximum R 0202 Current 0 01 Value R W 0300 Minimum R 0301 Maximum R 0302 Measured value 0 1 R 0307 Cu...

Page 20: ...bout parameters that you can change bit mask 0 bit 1 the device supports this option R 0703 1st bit Frequency 2nd bit Duration 3rd bit Current Lock status bit mask 0 bit Reserve R 0800 1st bit Interlo...

Page 21: ...into CW mode or set not zero frequency value to switch the device into QCW long pulses mode If you try to set a value more or less than limits then the value will be rounded to limit Any attempts to...

Page 22: ...cy value Setting a new frequency value frequency value 30 5 Some commands have a full and shortened version The full version is shown in the table and the shortened version is shown by UPPERCASE For e...

Page 23: ...om 0 00 Lower limit MAX Absolute maximum MIN Absolute minimum CURRent REAL Measured value TEC current 0 1 A UPper 0 0 Upper limit MAX Absolute maximum VOLTage REAL Measured value TEC voltage 0 1 V UPp...

Page 24: ...Maiman Electronics MBL1500B 23 Mechanical dimensions 18 All dimensions are in millimeters...

Reviews: