Maiman Electronics
MBL1500B
5
LD temperature protection window can be set in
Settings
submenu (paragraph 13.7) using parameters
TEC max temperature
and
TEC min temperature
.
Over-temperature protection
The MBL1500 has automatic overheating protection. If the set
TEC max temperature
is
exceeded, the LD power supply will be automatically turned off. After the temperature returns to
normal, the LD power supply can be switched on again.
State after switching on
After turning on the MBL1500 with the
Standby
button the LD power supply and TEC will always
be turned off.
Memorization of parameters
In case of an interruption/unplanned shutdown, the MBL1500 remembers the settings that were
set at the time of the last On/Off Laser or TEC command.