![Maico touchTymp MI 26 Operation Manual Download Page 77](http://html1.mh-extra.com/html/maico/touchtymp-mi-26/touchtymp-mi-26_operation-manual_3686551077.webp)
Operation Manual touchTymp MI 26 and MI 36 Version
8511780 Rev. 4
76
30/06/2021
5.6.19 Settings
– License Management – General
The License Management screen allows additional feature/test operation to be
incorporated into a base model by entering a license key. Contact MAICO or your local
distributor for more information.
Figure 169
The tab
General
contains a field to enter a new license
code in order to activate the license on the device. In the
middle, all available licenses are shown. The checkboxes
are activated automatically as soon as a license is
activated (Figure 169).
Figure 170
To enter a new license code, activate the keyboard by
pressing into the field
License code
and type in the code
Figure 171
If the code entered is invalid a message box will be shown
telling you to verify the code (Figure 171).
Ask your local distributor if any problems occur. If you
entered a correct code a message box will tell you
“Licensing completed
”.
5.6.20 Settings
– Service – General
Figure 172
Calibration reminder
: Annual calibration of the
touchTymp and its transducers is recommended.
Select or deselect this item to enable or disable a
reminder that will display daily. The reminder starts
1 month prior to the expiration of the calibration date for
your acoustic transducer(s) (Figure 172).
The user can always bypass the reminder message and
continue with screening.
Figure 173
Display calibration date
: Only for service (Figure 173).
Figure 174
Device calibration:
Only for service (Figure 174).
Figure 175
Restart to service mode:
Only for MAICO Technical
customer support (Figure 175).
Figure 176
Screen calibration:
Only for Service (Figure 176).
Figure 177
Test printout:
prints a test printout (without a session
result, Figure 177).