Operation Manual touchTymp MI 26 and MI 36 Version
8511780 Rev. 4
17
30/06/2021
3.3.4 Cleaning the Probe Tip
Also, check-out our training videos:
MAICO Training | touchTymp Part 1 | 7/8 Cleaning the probe - YouTube
https://www.youtube.com/watch?v=r7Wj5wDFQDg&list=PLonI5JzuDcd7lxKobEy7BW3DS59QCsbA
G&index=7
In order to secure correct immittance measurements it is important to make sure that the
probe system is kept clean at all times. Therefore please clean the probe on a periodic
basis. It is indispensable to remove cerumen from the probe tip’s small acoustic and air
pressure channels. Therefore please follow the illustrated instructions below. The pictures
show the procedure on the Pen Probe (left) and the Shoulder Box (right).
Never clean the probe tip while the tip is still attached to
the probe
Figure 2
Unscrew the probe cap by turning it in a counter clockwise
direction (Figure 2).
Figure 3
Take the plastic probe tip out of the probe (Figure 3).
Figure 4
Insert the blue end of the floss from back to front through
one of the probe channels. Pull the floss along its entire
length through the channel (Figure 4).
Figure 5
Proceed in the same way with all 4 probe channels. Use
the floss only once (Figure 5).
CAUTION