Operating Instructions MA 50
Geba_MA50_Fe_11a.doc
17 851 406/2 11/11
7 Computer interface
The MA 50 has a built-in serial RS 232 PC-interface for data transfer to a
connected PC. The serial cable to connect the MA 50 with the PC should be
a normal, non twisted 9-pin extension cable. It has a male and a female
connector and all wires are direct 1:1 connected. It is also sold as monitor
extension cable.
Caution: The computer you are using must meet electrical safety
requirements, such as IEC 950 or UL. This is to avoid electrical shock of either
the patient or you.
All test results are online transferred to the PC and stored in the database.
-
In pure tone audiometry initially
all measured data is transferred as
hearing threshold data. This is
indicated by the dot left from the
center of the LED-display
(see
picture 17).
-
To test the uncomfortable level
(UCL) press the test channel selector
key
longer than 2 seconds and the dot moves to the center of the LED-
display
. All measured data is now transferred as UCL data.
-
To test the most comfortable
level (MCL) press the test channel
selector button
longer than 2
seconds and the dot moves to a
position right of the center of the LED-display
. All measured data is now
transferred as MCL data.
Picture 17 MA 50 display with
pure tone threshold setting
Picture 18 MA 50 display with UCL setting
Picture 19 MA 50 display with MCL setting