Service Manual MA 42
D-0125965 Rev. D
25
18/05/2022
9. Under the Test Tab the following tabs can be selected: Attenuator Test, Patient
response Test, Diode Test as well as setting the Real time clock under the Service
Tab.
For the Attenuator linearity check select the Attenuator Tab.
Select the Tone and Tone type
Select the Input and a Level
Chose a Monitor Output and a
Monitor Level
10. Select the Patient Response Tab to check the functionality of Patient response
button.
11. Select Diode test Tab to test the functionality of all the diodes on the device.