Operating Instructions MA 25 / MA 25e
operating instructions_8012238-4_MA25_E_13a
16
8012238-4 10/13
7 Care and maintenance
Disconnect the power plug before cleaning!
If the surface of the instrument or parts of it is contaminated, it can be
cleaned using a soft cloth moistened with a mild solution of water and dish
washing cleaner or similar. The use of organic solvents and aromatic oils must
be avoided.
After each examination of a patient, it should be ensured that there is no
contamination on the parts in connection with the patient. General
precautions must be observed in order to avoid that disease from one patient
is conducted to others. If ear cushions are contaminated, it is strongly
recommended to remove them from the transducer before they are cleaned.
By frequent cleaning water should be used, but by severe contamination it
may be necessary to use a disinfectant. The use of organic solvents and
aromatic oils must be avoided.
Always disconnect the mains conductor during the cleaning process, and be
careful that no fluid is entering the inside of the instrument or the accessories;
no alcohol or spirits should be used.
8 Disinfection
It is recommended that parts which are in direct contact with the patient (e.g.
earphone cushions or patient response switch) are subjected to standard
disinfecting procedure between patients. This includes physically cleaning and
use of a recognized disinfectant. Individual manufacturer's instruction should
be followed for use of this disinfecting agent to provide an appropriated level
of cleanliness. If ear cushions are contaminated, it is strongly recommended to
remove them from the transducer before they are cleaned
Summary of Contents for MA 25e
Page 1: ...Operating Instructions MA 25 MA 25e...
Page 2: ......