19/10/2017
Table 15 Reference Values for Stimulus Calibration
REFERENCE VALUES FOR STIMULUS CALIBRATION
Fre-
quency
[Hz]
Reference equivalent threshold sound pressure level [RETSPL, dB re. 20
Pa]
according to ISO 389-1 with coupler IEC 60318-3
DD45
DD65
HOLMCO 8103
125
48.0
35.5
45.0
250
28.0
25.0
25.5
500
13.0
12.5
11.5
750
6.5
8.0
7.5
1000
6.0
7.0
7.0
1500
7.5
9.0
6.5
2000
7.5
5.0
9.0
3000
8.0
4.5
10.0
4000
7.0
7.8
9.5
6000
20.0
20.0
15.5
8000
11.0
12.0
13.0
Table 16 Frequencies and Maximum Intensities: AC (Air Condition) dBHL
TRANSDUCER MAXIMUM HEARING LEVELS
Frequency
[Hz]
Intensities [dB HL]
DD45
DD65
HOLMCO 8103
Tone
Tone
Tone
125
70
65
70
250
90
75
90
500
100
90
100
750
100
95
100
1000
100
95
100
1500
100
95
100
2000
100
95
100
3000
100
95
100
4000
100
95
100
6000
100
90
90
8000
90
85
80