5
8100216-3
(D-0100192)
3, clause 16. Any equipment not complying with the leakage current requirements in
IEC 60601-1 shall be kept outside the patient environment i.e. at least 1.5 m from the
patient support or shall be supplied via a separation transformer to reduce the leakage
currents. Any person who connects external equipment to signal input, signal output or
other connectors has formed a Medical Electrical System and is therefore responsible
for the system to comply with the requirements. If in doubt, contact qualified medical
technician or your local representative.
3.2.
Unpacking and Checking the MA 1
Checking for packaging and content damage
Thoroughly inspect the exterior of the shipping box for any sign of damage or
tampering. Should any damage be noted, please notify the carrier immediately. If the
content box has been damaged during transportation, the instrument should be
checked for any electrical or mechanical defects. Should any defects be identified,
please contact the responsible dealer. Keep all original packaging to facilitate any
insurance claims against the damages.
PLEASE KEEP ALL ORIGINAL PACKAGING FOR FUTURE USE!
The MA 1 is packaged in a specially-designed box. Please keep the box as it will be
useful for sending the instrument for the annual instrument check-up.
Please contact your nearest responsible dealer should the annual instrument check-up
be needed.
3.3.
Standard Accessories
DD 45 Audiometric Headset with HB7 Headband
Operation Manual
Audiogram Pad
Carry Case
2x AA Batteries
MA 1 Stand