Operating Instruction
ERO•SCAN
®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
43
8106556-5 02/15
18
Advanced Options for DPOAE Testing
NOTE: The Advanced Options for DPOAE Testing section is reserved for those units
purchased as a diagnostic device. Instruments purchased as a screener can skip to Section
20.
The Advanced Options menus permit modification of the test parameters and pass criteria
for the customizable DP protocols. Changes to the protocol should be made only by
qualified personnel, usually the administrator. If you are not familiar with the use of these
variables, do not attempt to change the protocols. Changes to any of these characteristics
may yield test results that differ from those obtained in other test modes.
The ERO•SCAN
®
instrument comes with pre-programmed protocol settings. See Appendix
D for the manufacturer settings of these protocols. Test protocol changes are saved in the
non-volatile memory so the settings will be retained even when the battery is discharged
temporarily.
18.1
Instructions for Customizing a Test Protocol
To enter the DPOAE Menu:
1)
Press
CHANGE at the main menu (Display 1, page 19).
2)
Using the
CHANGE
buttons, select the DPOAE protocol you want to
customize (the “DP 4s” protocol is not customizable).
3)
Press
SETUP at the Protocol menu (Display 2, page 19).
4)
At the Clock menu (Display 13, page 37) the teal arrow represents a
custom protocol menu is available. Hold down
CHANGE key for 3
seconds until the ‘READY’ light (green LED) turns off.
5)
At the New BT Device menu (Display 16, page 38) the teal arrow
represents a custom protocol menu is available. Hold down the
CHANGE key for 3 seconds until the ‘READY’ light (green LED) turns off.
You will now see the Level L1 screen (Display 28, page 44). You are in the DPOAE menu
and will be able to scroll the available protocol parameters with the
NEXT button and
make changes by using the
LEFT or RIGHT
arrow keys to CHANGE the selection.
NOTE: If you push the
DOWN arrow key without holding it for 3 seconds, you will scroll
through date and time, etc., rather than accessing the custom protocol menus.