Magtrol AN 2000 C
Instruction Manual
The figure 23.1 shows the indication (
FLt-E
) corresponding to entry stage of the
damping filter menu. Press the
ENTER
key to accede this menu.
ENTER
To access to program the filter level.
To skip over this menu and go to next one.
ESC
To exit from the programming mode without saving changes.
The figure 23.2 shows the initially selected level for the filter-E (any number between
0 and 9) with the FLT LED activated.
Press repeatedly the
key to change the digit until desired value appears on the
display.
ENTER
To save the entry into the memory and go to the next programming menu.
ESC
To exit from the programming mode without saving changes.
MENU 2B - DAMPING FILTER
The damping filter cuts off input values exceeding from the limits of a symmetrical band. This band becomes more selective as the
filter level is increased.
The filtering level is programmable from 0 to 9. Level 0 disables the filter.
[23.1] Damping filter
[23.2] Filter value