41
57-645 Eclipse Guided Wave Radar Transmitter - PROFIBUS PA
™
7.3
Specifications
7.3.1 Functional
System Design
Measurement Principle
Guided time-of-flight via time domain reflectometry
Input
Measured Variable
Level, determined by the time-of-flight of a guided radar pulse from
transmitter to product surface and back
Zero and Span
6 inches to 75 feet (15 to 2286 cm)
User Interface
Keypad
3-button menu-driven data entry and system security
Indication
2-line
×
8-character display
Digital Communication
PROFIBUS PA
™
Profile Version 3.0, MBP (31.25 kbits/sec)
Minimum Operating Voltage
9 VDC
Quiescent Current Draw
15 mA
DEV Revision
0X01
Function Blocks
AI_1, AI_2, AI_3, AI_4
Damping
Adjustable 0-10 seconds
Power
(Measured at instrument terminals)
General Purpose/Explosion Proof
9 to 32 VDC (15 mA maximum current draw)
IS/FISCO/FNICO
9–32 VDC (15 mA maximum current draw)
Housing
Material
Aluminum A356T6 (<0.20% copper), optional 316 stainless steel
Cable Entry
3
⁄
4
" NPT and M20
Environment
Operating Temperature
-40 to +175° F (-40 to +80° C)
Display Function Operating Temperature
-5 to +160° F
(-20 to +70° C)
Storage Temperature
-50 to +175° F (-40 to +80° C)
Humidity
0-99%, non-condensing
Electromagnetic Compatibility
Meets CE Requirements: EN 50081-2, EN 50082-2
Note: Twin Rod and Single Rod probes must be used in metallic vessel or
stillwell to maintain CE requirement.
Mounting Affects:
Twin Rod: Active rod must be mounted at least 1" (25 mm) from any surface or
obstruction. Minimum stillwell diameter for Twin Rod probe is 3".
Single Rod: Nozzles do not restrict performance by ensuring the following:
No nozzle is <2" (50 mm) diameter
Ratio of Diameter: Length is 1:1 or greater;
any ratio <1:1 (e.g., a 2"
×
6" nozzle = 1:3) may require a Blocking Distance
and/or DIELECTRIC adjustment
No pipe reducers are used
Obstructions
Keep conductive objects away from probe to ensure proper performance
Shock Class
ANSI/ISA-S71.03 Class SA1
Vibration Class
ANSI/ISA-S71.03 Class VC2