PRECAUTIONS
Inspect lamp before each use. If damage to the cord or housing is present, it is strongly
recommended the lamp be sent to a Magnaflux Authorized Service Center for repair.
CAUTION
Ultraviolet Radiation Source. UV-A Risk Group 2 (IEC/EN 62471).
CAUTION
Wear UV absorbing glasses at all times. Do not use at less than minimum
working distance.
WARNING
Do not immerse lamp in any liquid. Connect only to grounded electrical outlets.
CAUTION
Do not abuse cords. Never carry lamp by cord or yank to disconnect from
receptacle. Never retrieve lamp by pulling on cord. Do not suspend the lamp
by the cord.
CAUTION
Only use outdoor rated extension cords SW-A, SOW-A, STW-A, STOW-A, SJW-
0A, SJOW-A, SJTW-A, or SJTOW-A. Cord Rating 0-6 A.
0-25 ft / 0-7.6 m: 18 AWG / 1.0 mm
2
26-100 ft / 7.6-30.4 m: 16 AWG / 1.5 mm
2
101-150 ft / 30.4-45.7 m: 14 AWG / 2.5 mm
2
OPERATION
1.
Connect the lamp to the power supply.
2.
Connect the power supply to line power.
3.
Press the power button on the rear of the lamp housing to turn on the UV LEDs.
4.
The lamp will immediately be at full intensity and will stabilize within 5 minutes.
5.
On EV6500 models only switch from UV mode to visible light mode, press and hold the power
button for 3 seconds.
6.
To turn off, press the power button on the rear of the lamp housing.
NOTE: The EV6000 and EV6500 are designed for continuous use; operating the lamp only
when required will increase the lifespan of the UV LEDs.
MAINTENANCE
•
Check UV-A intensity at regular intervals using a calibrated UV-A Meter (320-400 nm range)
•
UV-A intensity readings should be taken with the lamp positioned 15 in / 38 cm from the
lamp filter to the face of the sensor
•
Use mild soap and a damp cloth to clean the lamp housing as needed
•
Do not spray lamp or power supply or immerse any part in liquid