A. Test the Tip of the W2.
1. Please put the Tip Holder Set (not provided in
accessary) into the W2 of T-Rack.
2. Please add 1 ml water to the Sample Tube and put it in
the W1 of the T-Rack.
3. Please select the Test W2 Tip.
4. The test can be divided into two stages. The first stage
is to observe the smooth of taking the tip out of the
Tip Holder. Please check whether the tip holder is
taken out together with the tip. If the Tip Holder is
taken out together with the tip, please reset the Tip
Holder and try it again.
5. Please check whether the liquid levels of all tips are
consistent with each other. If the liquid levels of all tips
are not consistent with each other, please check the
O-Ring
(Please refer to the Replace O-Ring chapter for more
information).
B. Test the Tip of the W3.
1. Please put the Pipette Tip into Well 3 of T-Rack.
2. Please add 1 ml water to the Sample Tube and put it in
the Well1 of the T-Rack.
3. Please select the Test W3 Tip.
4. Please check whether the liquid levels in all tips are
consistent with each other. If the liquid levels are not
consistent, please check the O-Ring
(Please refer to the
Replace O-Ring chapter for more information).
C. Test the Tip of the W4.
(MagCore Super Only)
1. Please put the 200μl SP Tip into Well 4 of T-Rack.
2. Please add 1 ml water to the Sample Tube and put it in
Well1 of the T-Rack.
3. Please select the Test W4 Tip.
4. Please check whether the liquid levels in all tips are
consistent with each other. If the liquid levels are not
consistent, please check the O-Ring
(Please refer to the
Replace O-Ring chapter for more information).
Piston Test
Test W3 Tip
Test W4 Tip
Test W2 Tip