MagCore Super Operation Manual Download Page 18

17

www.rbcbioscience.com

A. Test the Tip of the W2.

1.  Please put the Tip Holder Set (not provided in 

accessary) into the W2 of T-Rack.

2.  Please add 1 ml water to the Sample Tube and put it in 

the W1 of the T-Rack.

3.  Please select the Test W2 Tip.

4.  The test can be divided into two stages.  The first stage 

is to observe the smooth of taking the tip out of the 

Tip Holder.  Please check whether the tip holder is 

taken out together with the tip.  If the Tip Holder is 

taken out together with the tip, please reset the Tip 

Holder and try it again.

5.  Please check whether the liquid levels of all tips are 

consistent with each other.  If the liquid levels of all tips 

are not consistent with each other, please check the 

O-Ring 

(Please refer to the Replace O-Ring chapter for more 

information).

B. Test the Tip of the W3.

1.  Please put the Pipette Tip into Well 3 of T-Rack.

2.  Please add 1 ml water to the Sample Tube and put it in 

the Well1 of the T-Rack.

3.  Please select the Test W3 Tip.

4.  Please check whether the liquid levels in all tips are 

consistent with each other.  If the liquid levels are not 

consistent, please check the O-Ring

 (Please refer to the 

Replace O-Ring chapter for more information).

C. Test the Tip of the W4. 

(MagCore Super Only)

1.  Please put the 200μl SP Tip into Well 4 of T-Rack.

2.  Please add 1 ml water to the Sample Tube and put it in 

Well1 of the T-Rack.

3.  Please select the Test W4 Tip.

4.  Please check whether the liquid levels in all tips are 

consistent with each other.  If the liquid levels are not 

consistent, please check the O-Ring 

(Please refer to the 

Replace O-Ring chapter for more information).

Piston Test 

Test W3 Tip

Test W4 Tip

Test W2 Tip

Back

Piston Test 

Test W3 Tip

Test W4 Tip

Test W2 Tip

Back

Piston Test 

Test W3 Tip

Test W4 Tip

Test W2 Tip

Back

Stop

Piston Test 

Next Step

Stop

Piston Test 

Next Step

Stop

Piston Test 

Next Step

Stop

Piston Test 

Next Step

Summary of Contents for Super

Page 1: ...UserInterface 10 EmergencyStop 10 StartPrograms 11 Iftheoptictestresultis 12 overdetectionlimit 12 Stop 14 Enablethebarcodefunction 14 Barcodescanningandediting 15 Optictestdata 15 Easymaintenance 16...

Page 2: ...of the type of power supply at your lab consultthesellerorlocalpowercompany Forproductsintendedtooperatefrom batteryorothersources refertotheoperatinginstructions 11 Grounding or Polarization This pro...

Page 3: ...edbytheClot Detectionfunctionoftheinstrument Clumpingofbeadsmayleadtolowyieldandpurityofnucleicacids cross contamination andinhibitionofdownstreamassays e g PCR Aftertherunhasfinished carefullyinspect...

Page 4: ...oesnotwarrantthatsamplestreatedwithLysisBufferarecompletelyinactivatedand noninfectious Aftersampleprocessingiscompleted removeandautoclavealldisposableplastics Donoteat drinkorsmokeinthelaboratorywor...

Page 5: ...ts Warning electricity Instructionmanualmustbereadbeforeoperating Consultinstructionsforuse Alternatecurrentinput Use byDate Nouseperiod Temperaturestorageconditions15 C 30 C Cataloguenumber productco...

Page 6: ...0xH770 mm W27 95xD29 92xH30 31 inches NetWeight 76 5kg 168 68LB MagCoreSuper 68 5kg 151LB MagCoreHF16Plus Specification Operating Parameters Operating Environment Applications Whole Blood DNA 200 l 40...

Page 7: ...vetteSet 10pcs X1 11 WasteBoxX1 12 BarcodeScannerX1 MagCoreSuperOnly Please refer to the machine size shown in the above figure to prepare an appropriate position for the machine and reservespaceforth...

Page 8: ...alde16slots n mero1 16daesquerdaparaa direita 12345678910111213141516 Printer MagCoreSuperOnly TouchPanel FrontDoor TopServiceLid CoolingFans BackServiceLid OperationArea USBPort ScannerPort BarcodeSc...

Page 9: ...etoputitintothewellifthe directionisnotcorrect 2 Putthe200 lSPtipintotheW4oftheT Rack Pleaseinsertthefrontendofthecartridgeintothespace belowthefixingplateoftheCartridgeRack 1 Putthetipintothecorrespo...

Page 10: ...3 Testheater PleaserefertoHeaterTestchapterfortheinformationabout theoperation Step4 Testopticmodule MagCore SuperOnly PleaserefertoOpticTestchapterfortheinformationabout theoperation Whentheopticmodu...

Page 11: ...e The above figureshowsthe function of the operation icons Start Maintenance Barcode Optics Data Please read the security information to correctly stop the machine whenanaccidenttakesplace When an eme...

Page 12: ...ystemwillaskyouwhetherornot todeletepreviousbarcoderecord Ifthereisno barcoderecord youwillnotbeasked MagCore Superonly HF16Plusoptional Start Maintenance Barcode Optics Data 401 ProgramCode SelectElu...

Page 13: ...rtthe USBFlashDriveandpresstheSavebutton tosavetocurrentmeasureddata After finishingtheabovesteps pleasepressNext Step 10 7 8 9 102 ProgramCode SelectOpticMeasurementFunction Enable Disable Back Press...

Page 14: ...disappears Ifthefilesavedlasttimeislost you canretrievethisfileaccordingtothe instructionsofoptictestdatachapter OK Next Step Back 10XDilution 1 Take away and store the products which are NO need to...

Page 15: ...esstheAbortbuttonifyouwanttoend thetest Duringtheabortionprocess moduleswillreturnto theiroriginalpositions Pleasedonotoperatethe machineduringtheprocess Themachinehasbeenstoppedandreset please presst...

Page 16: ...code Optics Data Barcode Cancel Scan Print Data Delete All OpticalData Display Optical Data Print Optical Data Back Display Previous Optical Data Start Optical Measurement Print Previous Optical Data...

Page 17: ...stonTest Easymaintenance 2 1 Start Maintenance Barcode Optics Data Stop O Ring Next Step Syringe O Rings 1 2 3 4 Maintenance Reject Tip Piston Test UV Heater Test Replace O Ring Motor Test Next Back M...

Page 18: ...Well3ofT Rack 2 Pleaseadd1mlwatertotheSampleTubeandputitin theWell1oftheT Rack 3 PleaseselecttheTestW3Tip 4 Pleasecheckwhethertheliquidlevelsinalltipsare consistentwitheachother Iftheliquidlevelsareno...

Page 19: ...ips are still on the pistons you have to press Reject Tip to remove thetipsfromthepistons PleaseselecttheRejectTipbutton Clean Piercing Needles 2 1 1 Stop PiercingNeedles Next Step Prev Maintenance Up...

Page 20: ...ptheD2lamp Thescreenwillshowthemeasuredvaluesafter thetestisfinished YoucaninserttheUSBFlash DriveandthenpresstheSavebuttontostorethe currentdataaftertheUSBiconshowsinthestatus bar PleasetakeouttheUSB...

Page 21: ...edles System Status Optic Test Alarm History Back Clear Save to USB AlarmMessage Back Date Time ErrorMessages Code AlarmMessage Back 0000001 ErrorCode Pleaserefertousermanualor contactyourlocalenginee...

Page 22: ...methatyouchangedin Current Time afterthechangeisfinished System Status Prev Maintenance Update Program Vender Piercing Needles System Status Optic Test Alarm History Back SystemStatus Back System Time...

Page 23: ...sfinished pleasepressOKtocheck whethertheUIversioniscorrect Pleasedumpthewastewhentheprogramis completed andwashthewasteboxeverytime Warning Pleaseconfirmthenameof UIversion Iftheversionisnot correct...

Page 24: ...following figure and thenclosethecoveroftheprinter If the position of the thermal paper roll is not correct the red yellow lamp will blink Please press the Feed button to confirm the installation is...

Page 25: ...pdated the following sections General Considerations Document Revision History Update to comply with the requirements of the IVD Regulation EU 2017 746 June 2023 EN MagCore Super HF16 Plus Operation M...

Page 26: ...15F No 15 QiaoheRd ZhongheDist NewTaipeiCity235029 Taiwan TEL 886289121200FAX 86289121300...

Reviews: