17 

Problem 

Potential cause 

Solution

 

The machine will not 
switch on 

A. Inadequate voltage or 
blown fuse 

B. Defective cable 

C. Regulator fault 

A. Check voltage and fuse 

B and C. Contact the 
Technical Services 
Department 

 

The machine will not 
commence cooling 
operations 

A. Set Point too high 

B. Fault in regulator 
parameters 

C. Refrigeration-circuit 
fault 

A. Lower Set Point 

B and C. Contact the 
Technical Services 
Department 

The machine will not 
commence heating 
operations 

 

A. Set Point too low 

B. Fault in regulator 
parameters 

C. Heating-circuit fault 

A. Raise Set Point 

B and C. Contact the 
Technical Services 
Department 

The machine’s 
compressor or external 
casing are too hot 

Machine inadequately 
ventilated, or exposed to 
sunlight or other heat 
source 

 

Move machine to another 
location 

 

Internal temperature of 
the machine and the 
temperature displayed 
on the regulator do not 
match 

Internal temperature of 
the machine calibrated 
incorrectly 

Contact the Technical 
Services Department 

 

 

 

 

 

 

 

 

 

Summary of Contents for Z-100

Page 1: ...Ver 04 Febrero 2018 C MARA DE CONSERVACI N PRESERVATION UNIT Z 100 MANUAL DE INSTRUCCIONES INSTRUCTION MANUAL...

Page 2: ......

Page 3: ...conectado desde la clavija Desenchufar el aparato antes de realizar trabajos de limpieza y mantenimiento no tirar del cable para desenchufarlo Si el acceso al enchufe es dif cil desconectar la corrien...

Page 4: ...regulador digital que controla la temperatura en el interior Posee un sistema de alarma que permite advertir al usuario de cualquier deviaci n peligrosa que pueda alterar la perfecta conservaci n del...

Page 5: ...able para la conexi n el ctrica situado junto al compresor en la parte trasera Importante Verifique que la tensi n de su instalaci n corresponde con la tensi n de la placa de identificaci n de su equi...

Page 6: ...ura de consigna Una vez que se conecta la c mara el regulador muestra constantemente la temperatura medida por la sonda ubicada en el interior De f brica el equipo viene programado y calibrado para tr...

Page 7: ...10 minutos se dispara el zumbador Por ejemplo si se tiene el equipo programado para mantener 16 C subir por encima de 17 5 C o bajar por debajo de 14 5 C durante m s de 10 minutos disparar la alarma...

Page 8: ...mo 200 blister 170 tubos 200 botes Resoluci n medida de temperatura 0 1 C Rango de temperatura Desde 15 C a 40 C 0 5 C Dimensiones exteriores 896x576x665 mm Peso en vac o 29 Kg Condiciones funcionamie...

Page 9: ...po facilitando as la evacuaci n de calor necesaria para una correcta producci n de fr o en el interior Normalmente a la temperatura de trabajo de 16 no se formar un exceso de escarcha en la pared fr a...

Page 10: ...P ngase en contacto con el Servicio T cnico La c mara de conservaci n no calienta A Set Point bajo B Fallo en par metros del regulador C Aver a circuito calefactor A Revise el Set Point B y C P ngase...

Page 11: ...enen otras vac as Distribuir la carga de manera apropiada para que el aire pueda recircular libremente Emplear siempre bandejas originales de alambre que permiten la recirculaci n del aire con total l...

Page 12: ...ry The electricity supply to the machine is only cut off when it has been unplugged at the socket Unplug the machine before cleaning it or carrying out any maintenance work Do not unplug it by pulling...

Page 13: ...hich controls the temperature inside the container It also has an alarm system which allows the operator to check whether a serious deviation in temperature has occurred at any time that would affect...

Page 14: ...power supply located at the rear next to the compressor Important Make sure the voltage of the electricity supply matches that indicated on the container s specification plate For the safety of the d...

Page 15: ...ermometer and displays these continuously on the screen The machine is programmed and calibrated at the factory to operate at a default temperature of 16 C However if you wish to assign a different te...

Page 16: ...ample if you have set the device to maintain a temperature of 16 C and the temperature either rises above 17 5 C or falls below 14 5 C for longer than ten minutes the buzzer will sound If the buzzer s...

Page 17: ...mum no of doses 200 blisters 170 tubes 200 jars Temperature measurement resolution 0 1 C Temperature range 15 C to 40 C 0 5 C External dimensions 896x576x665 mm Unladen weight 29 kg Operating conditio...

Page 18: ...d with heat evacuation which is necessary to ensure the interior of the machine can be kept suitably cool Normally at a working temperature of 16 C excessive amounts of frost will not form on the mach...

Page 19: ...echnical Services Department The machine will not commence heating operations A Set Point too low B Fault in regulator parameters C Heating circuit fault A Raise Set Point B and C Contact the Technica...

Page 20: ...mpty Distribute the doses inside the machine in such a way that the air can circulate freely Always use the wire shelves that are supplied with the machine as these allow the air to circulate without...

Page 21: ...19...

Page 22: ...20...

Page 23: ...21...

Page 24: ...22 Pol Ind Valdeferr n Calle 5 50600 Ejea de los Caballeros Zaragoza ESPA A Tel 34 976 66 29 14 Fax 34 976 67 18 88 E mail magapor magapor com www magapor com...

Reviews: