background image

Operating Instructions

Powering ON

Ÿ

Ensure the battery is fresh (voltage above 7.2v)

Ÿ

Set the power switch to the ON Position

. MAG-ic probe can also be powered via the USB cable 

to your PC. However, computers differ and cannot guarantee that the 5V standard is exact. The 
initial calibration will take care of small differences in USB power but may not be as accurate as 
when using battery power.

Ÿ

Press Function Button

Ÿ

Attach magnet to Probe

. MAG-ic probe is supplied with 2 magnets, a ½” ball and a smaller 

magnet. Your first choice should always be the ½” ball and only use the smaller magnet when 
access is limited by small f-holes etc.

Ÿ

Press Function button and 

Remove Magnet

 at least 6 inches from probe when prompted.

Ÿ

If 

Calibration is Successful

, you will be prompted to re-attach the magnet and press function key.

This completes the initial calibration and MAG-ic Probe is ready to perform your measurements. The 
display should read 0.00 when the magnet is attached to the probe. If the advanced calibration is 
required, perform it now - (see Advanced Calibration)

HOLD Function

MAG-ic Probe features a 

HOLD

 mode to assist you in obtaining accurate measurements. There are 2 

ways to invoke the 

HOLD 

feature. Either slide the power switch to the 

HOLD

 position, OR, press 

BUTTON 3. The 

HOLD

 mode is indicated by an 

“H”

 preceding the mils display. A HOLD mode can also 

be activated in the MAG-ic Probe Software by clicking on the HOLD/FREE icon.

The 

HOLD

 mode can be used at every measuring location, to ensure that the lowest reading is being 

displayed. The 

HOLD

 mode will freeze the display and only allow lower readings to update the display. 

Release the hold mode to continue normal display reading.

Summary of Contents for V4.0

Page 1: ...lds are strong at the magnet surface and deteriorate rapidly in a non linear manner as distance increases Ÿ Magnetic fields will vary with changing temperatures Ÿ Magnetic and most other sensors are unstable and vary their output with changing temperatures Ÿ Friction may cause the inner magnet to not be centered on the probe tip when moving the probe across the measured object Considering the abov...

Page 2: ...be engaged by pressing FUNCTION 1 Firstly the initial calibration procedure will compensate for differences between your shop environment and the factory environment where MAG ic Probe was calibrated The problem is that after this calibration your hand may heat the probe causing a small drift in value This is the reason for the 2nd calibration as described above To assist in this 2nd calibration t...

Page 3: ...rts currently installed on your pc Look for the USB Serial Port COM and note the port number If you double click this line you should see a similar properties window as below right In the above example COM 3 is the port that will be used when we connect to the MAG ic Probe Software If a driver is not automatically installed Please return to the website and click on Document and Driver Downloads Do...

Page 4: ...utton to see the available ports The Serial port should be identified as something like USBmodem or USBSerial etc Pick this option and then click on the small box to the right of this line This box will provide the full detail of the Port Name In this example it is Usbmodem411 but each computer will be different The full name is dev cu usbmodem411 Make a note of this full name as you will need it ...

Page 5: ...ssful you will be prompted to re attach the magnet and press function key This completes the initial calibration and MAG ic Probe is ready to perform your measurements The display should read 0 00 when the magnet is attached to the probe If the advanced calibration is required perform it now see Advanced Calibration HOLD Function MAG ic Probe features a HOLD mode to assist you in obtaining accurat...

Page 6: ...uring instrument you want to equalize the 2 instruments by adding or subtracting a value The OFFSET value is set in mils 1 1000th inch Please remember to set the offset back to zero when completed for normal measures Press 2 for OFFSET followed by either 2 for negative decrement or 4 for positive increment Re Calibration Press 4 for Re calibration This will restart the complete initial calibration...

Page 7: ...unit when not in use to conserve battery life Ÿ Take care of the magnets they are fragile and will shatter or splinter if allowed to snap onto each other or metal objects Ÿ Ensure your instrument does not contain any metallic objects to which the internal magnets can be attracted Removing a magnet from a finished instrument can be difficult Ÿ Do not drag the probe across finished surfaces Either l...

Page 8: ......

Reviews: