Electrochemical cells calibration
MaMoS gas analyser
I. GAS CALIBRATION OF ELECTROCHEMICAL CELLS
1. Turn the analyser ON and connect it using proper connection interface to the PC computer.
2. Start the service program MaMoS on the connected PC computer and communicate with the analyser.
3. Press F6 and enter password:
010939
4. Make sure that:
-the analyser is in the measurements phase (user's manual chapter
7.3.8. Work modes
)
-time of the measurement cycle is long enough to prevent switching the analyser during calibration to
ventilation mode.
-analyser is sucking fresh air
5. Open Service->CPU from the top menu.
3