Stopping the test
1. If the patient is troubled by the test, press the
Start/Stop
button on the
device or the
Start/Stop
button on the probe control pad on the
shoulder strap.
2. The test is interrupted and the pump pressure is relieved immediately.
Already measured results are kept when a measurement is stopped.
Making additional tympanograms
1. To make a new sweep, press the
Start/Stop
button.
If you wish to overwrite a tympanogram, use the dial or the soft-buttons
Previous Curve
or
Next Curve
to select the
curve you wish to overwrite.
When you have finished testing one ear, switch ears, if needed.
Viewing and managing test results
2. To view tympanometry results, see
.
3. When you have finished testing, see
Managing test results in MADSEN Zodiac
9.3.1
The Tympanometry screen
A.
On-line values
B.
Tympanometry graph
C.
Reflex Screening results
D.
Tympanometry results
E.
On-screen selections
Otometrics - MADSEN Zodiac
55
9 Testing with Zodiac Diagnostic or Clinical - Stand-Alone
Summary of Contents for zodiac
Page 1: ...MADSEN Zodiac Reference Manual Doc No 7 50 1610 EN 02 Part No 7 50 16100 EN...
Page 21: ...4 2 Bottom view A Probe connections B Label Otometrics MADSEN Zodiac 21 4 Views of Zodiac...
Page 156: ...jB jB App 2 5 2 G App 2 5 3 B G B G Y B G B G 156 Otometrics MADSEN Zodiac App 2 MADSEN Zodiac...
Page 157: ...App 2 5 4 226 Hz 226 Hz 200 daPa Otometrics MADSEN Zodiac 157 App 2 MADSEN Zodiac...
Page 206: ...152 151 151 152 150 152 206 Otometrics MADSEN Zodiac Index...