
Product Notes
How it Works
Once the logger is started samples are taken on each axis (X, Y and Z) every 1024Hz
(1 ms). These samples are compared with the trigger values and if it crosses the threshold
value then it records the timestamp and starts logging the triggered data.
If the samples are not above the threshold value then it is logged as the pre-trigger data.
Since only 50 pre-trigger values can be recorded the data gets overwritten if the logger is
not triggered within 50 pre-trigger values.
The sampling frequency is the same for all reading rates therefore the samples are skipped
based on the reading rate. For example: if the reading rate is 512Hz the samples are taken
every 1ms but every other sample gets logged instead of logging every sample.
Trigger Settings
Below is a table of default trigger values based on the range of the logger. All axis are
enabled to check for triggers at these g levels.
To edit trigger values, select
Start Device
from the
Start Menu
. Click
Yes
, then select
Trigger Settings
. Click the
Change
button. Each axis can be set to trigger at a certain g
force level. Check the
Enable Trigger
box to enable the trigger for that axis. The values
can be typed in manually or configured by using the slider control. Click
Save
to save the
changes.
To initiate the TSR101 to constantly sample, set the high and low trigger settings to 0 g.
Axis Orientation
When the TSR101 is laying flat, with the label facing
up, the x-axis indicates left to right direction and will
read ~0 g when still. The y-axis reads in a top to bottom
direction and reads ~0 g when still. The z-axis reads
perpendicular to the other two axis, and will read ~1 g
when still.
Installation Guide
Installing the Interface cable
- IFC200
Insert the device into a USB port. The drivers will install automatically.
Installing the software
Insert the Software USB Stick in an open USB port. If the autorun does not appear, locate
the drive on the computer and double click on
Autorun.exe
. Follow the instructions
provided in the Wizard.
TSR101
Product Information Card
Device Operation
Connecting and Starting the data logger
- Once the software is installed and running, plug the interface cable into the data logger.
- Connect the USB end of the interface cable into an open USB port on the computer.
- The device will appear in the Connected Devices list, highlight the desired data logger.
- For most applications, select “
Custom Start
” from the menu bar and choose the desired
start method, reading rate and other parameters appropriate for the data logging
application and click “
Start
”. (
“
Quick Start
” applies the most recent custom start options,
“
Batch Start
” is used for managing multiple loggers at once, “
Real Time Start
” stores the
dataset as it records while connected to the logger.
)
- The status of the device will change to “
Running
”, “
Waiting to Start
” or “
Waiting to
Manual Start
”, depending upon your start method.
- Disconnect the data logger from the interface cable and place it in the environment to
measure.
Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device
cannot be restarted until it has been re-armed by the computer.
Downloading
data
from a
data
logger
- Highlight the data logger in the Connected Devices list. Click “
Stop
” on the menu bar.
- Once the data logger is stopped, with the logger highlighted, click “
Download
”. You will
be prompted to name your report.
- Downloading will offload and save all the recorded data to the PC.
Device Maintenance
Battery Replacement
Materials:
3/32” HEX Driver (Allen Key)
Replacement Battery (U9VL-J)
- Remove the cover from the device by unscrewing the four screws.
- Remove the battery from its compartment and unsnap it from the connector.
- Snap the new battery into the terminals and verify it is secure.
- Replace the cover taking care not to pinch the wires. Screw the enclosure back together
securely.
Note: Be sure not to over tighten the screws or strip the threads.
Recalibration
The TSR101 standard calibration is performed at 0 g for the X-axis, 0 g for the Y-axis and
1 g for the Z-axis.
Pricing:
Recalibration with Certificate (not NIST traceable) $76.00
Recalibration
$46.00
Prices and specifications subject to change. See MadgeTech’s terms and conditions at www.madgetech.com.
To send the devices back, visit www.madgetech.com, select Services then RMA Process.
Range
5 g
50 g
100 g
250 g
Trigger Value
2.5 g
25 g
52.5 g
120 g
Y-Axis
Z-Axis
X-Axis