MadgeTech Temp101 Product Information Card Download Page 1

MadgeTech, Inc.

PO Box 50 

l

 

Warner, NH 03278

Phone 603.456.2011

 

l

 

Fax 603.456.2012

www.madgetech.com

 

l

 

[email protected]

Red 485
Dark Blue Pantone 289
Light Blue 58% of Pantone 289

REV F 2009.04.30

To view the full MadgeTech product line,

visit our website at 

www.madgetech.com.

Temp110

Miniature Temperature Recorder with 10 Year Battery Life

Red 485
Dark Blue Pantone 289
Light Blue 58% of Pantone 289

Product Information Card

Temp100, Temp101, Temp110 and

TempRetriever

Temp101

Miniature Temperature Recorder

Temp100

Miniature Temperature Recorder with PushButton Start/Stop

TempRetriever

Miniature Temperature Recorder with PushButton Start/Stop

Part Number

Temp100

Temp101

Temp110

TempRetriever

Sensor

Internal semiconductor

Range

–40 to +80°C

Resolution

0.1°C

Accuracy

+0.5°C (0 to +50°C)

Memory

32,767

32,767

32,767

16,383

Sample Rate

2 seconds up to 

12 hours

2 seconds up to 

12 hours

2 seconds up to 

12 hours

5 seconds up to 

30 minutes

LED Indicator

Red & Green

Red

Red

Red & Green

Pushbutton Start

Yes

No

No

Yes

Alarm

Yes

Yes

No

Yes

Req. Interface Package

IFC110 or IFC200

Baud Rate

2,400

2,400

57,600

38,400

Battery Life

1 year

1 year

10 years

1 year

Operating Environment

–40 to +80°C, 0 to 95%RH  non–condensing

Material

ABS plastic

Dimensions

1.4” x 2.2” x 0.6” 

(36mm x 56mm x 

16mm)

1.4” x 2.2” x 0.6” 

(36mm x 56mm x 

16mm)

1.7” x 2.3” x 0.8” 

(44mm x 59mm x 

21mm)

1.4” x 2.2” x 0.6” 

(36mm x 56mm x 

16mm)

Approvals

CE

Product Quick Reference Card

Specifications subject to change.

See MadgeTech’s terms and conditions at www.madgetech.com

Device Maintenance

Battery Replacement

Materials: 

Small Phillips Head Screwdriver and LTC-7PN battery 

Temp101 and Temp110:

- Puncture the center of the back label with the screw driver 

and unscrew the enclosure.

- Remove the battery by pulling it perpendicular to the board.

- Insert the new battery into the terminals and verify it is secure.

- Screw the enclosure back together securely.

 

Temp100 and TempRetriever:

- Puncture the center of the back label with the screw driver 

and unscrew the enclosure.

- Use a small non-metallic device to push the coin-cell battery 

out of its holder.

- Hold down the pushbutton in the middle of the circuit board 

and insert the new battery into the holder.

- Release the pushbutton and watch to verify that the LED’s 

blink. (If they do not, then you may have inserted the battery 

upside down.  Remove the battery and try again.)

- Replace the circuit board in the enclosure.

- Screw the enclosure back together securely.

Battery Warning

WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD.  DO NOT 

SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, DISASSEMBLE, 

CRUSH,  PENETRATE  OR  INCINERATE.  BATTERY  MAY  LEAK  OR 

EXPLODE IF HEATED ABOVE 80°C (176°F).

Recalibration 

Standard calibration is one point at 25°C for all 4 devices.

Pricing:

Recalibration traceable to NIST  $60.00

Recalibration 

$40.00

 

Additional:

As Found Data 

$15.00 per parameter/channel

Verification Point 

$15.00 per point

To  send  the  devices  back,  visit  www.madgetech.com,  select  

Services

 then 

RMA Process

.

- Gently pry the two halves of the enclosure apart.

- Remove the circuit board.

- Use a small non-metallic device to push the coin-cell battery out of its 

holder.

- Hold down the pushbutton in the middle of the circuit board and insert 

the new battery into the holder.

- Release the pushbutton and watch to verify that the LED’s blink. (If they 

do not, then you may have inserted the battery upside down.  Remove 

the battery and try again.)

- Replace the circuit board in the enclosure.

- Press the two halves of the enclosure back together.