SciTemp140-
PT
Device Operation (cont’d)
Trigger Settings
The device can be programmed to only record
based off user configured trigger settings.
1.
In the
Connected Devices
panel, click the
device desired.
2.
On the Device Tab, in the
Information
Group, click
Properties
. Or, right-click the
device and select
Properties
in the context
menu.
3.
Select
Trigger
in the Properties window.
4.
Trigger formats are available in
Window
or
Two Point Mode
. Window mode allows a
high and/or low trigger set point, and a
trigger sample count or “window” of time
recorded when set points are exceeded to
be defined. Two point allows for different
Start and Stop setpoints to be defined for
both the high and low triggers.
Set Password
To password protect the device so that others
cannot start, stop or reset the device:
1.
In the Connected Devices panel, click the
device desired.
2.
On the Device Tab, in the Information
Group, click Properties. Or, right-click the
device and select Properties in the context
menu.
3.
On the General Tab, click Set Password.
4.
Enter and confirm the password in the box
that appears, then select OK.
Device Maintenance
O-Rings
O-ring maintenance is a key factor when
properly caring for the SciTemp140-PT. The
O-rings ensure a tight seal and prevent liquid
from entering the inside of the
device.
Battery Replacement
Materials:
SD900097-00
1.
Unscrew the bottom of the logger and
remove the battery.
2.
Place the new battery into the logger.
Note the polarity of the battery. It is
important to insert the battery with
positive polarity pointing upward
towards the probe. Failure to do so
could result in product inoperability or
potential explosion if exposed to high
temperatures.
3.
Screw the cover back onto the logger.
Recalibration
Scigiene recommends annual recalibration. To
send devices back for calibration, contact
Scigiene.
Note: This product is rated for use up to 140 °C (284 °F). Please heed the battery warning. The
product will explode if exposed to temperatures above 140 °C (284 °F).