background image

Wiring the Data Logger

Wiring Options

For 4-wire RTD probes, connect the four lead wires to your RTD logger as shown in the 
figures below.

For 3-wire RTD probes, short inputs 3 
and 4 together, then connect the lead 
wires to inputs 1, 2 and 3.
For 2-wire RTD probes, short inputs 
3 and 4 together and inputs 1 and 2 
together, then connect the RTD lead 
wires to inputs 2 and 3.

 
Warning: Note the polarity instructions. Do not attach wires to the wrong terminals. 100 

, 2 or 4 wire RTD probes are 

recommended for the most accurate performance. Most 100 

, 3 wire RTD probes will work, but MadgeTech cannot guarantee the 

accuracy. To determine whether or not the 3-wire RTD probe will work, the resistance between the two same colored wires should be 
less than 1 

. (Note: Please contact the manufacturer of the RTD probe for questions on the resistance)

 

Installation Guide

Installing the Interface cable

- IFC200
  Insert the device into a USB port. The drivers will install automatically.
- USB-1 or USB-101
  Install the USB drivers from the CD provided in the kit, then plug the USB cable into the 

computer and the serial cable into the serial port.

Installing the software

The Software can be downloaded from the MadgeTech website at the following link: www.
madgetech.com/software-download. Follow the instructions provided in the Installation
Wizard. 

Device Operation

Connecting and starting the data logger

- Once the software is installed and running, plug the interface cable into the data logger.
- Connect the USB end of the interface cable into an open USB port on the computer. 
- The device will appear in the

 

Connected Devices

 

list, highlight the desired data logger.  

- For most applications, select “

Custom Start

” from the menu bar and choose the desired 

start method, reading rate and other parameters appropriate for the data logging 
application and click “

Start

”. (

Quick Start

” applies the most recent custom start options, 

Batch Start

” is used for managing multiple loggers at once, “

Real Time Start

” stores the 

QuadRTD and OctRTD

Product User Guide

dataset as it records while connected to the logger.)

- The status of the device will change to “

Running

”, “

Waiting to Start

” or “

Waiting to 

Manual Start

”, depending upon your start method.

- Disconnect the data logger from the interface cable and place it in the environment to 

measure.

 
Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device 
cannot be restarted until it has been re-armed by the computer. 

Downloading data from a data logger

- Connect the logger to the interface cable.
- Highlight the data logger in the Connected Devices list. Click “

Stop

” on the menu bar.

- Once the data logger is stopped, with the logger highlighted, click “

Download

”. You will 

be prompted to name your report.

- Downloading will offload and save all the recorded data to the PC. 

Device Maintenance

Battery Replacement

Materials

:

3/32” HEX Driver (Allen Key)
Replacement Battery (U9VL-J)

- Remove the cover from the device by unscrewing the four screws.
- Remove the battery from its compartment and unsnap it from the connector.
- Snap the new battery into the terminals and verify it is secure.
- Replace the cover taking care not to pinch the wires. Screw the enclosure back together 

securely.

 
Note: Be sure not to over tighten the screws or strip the threads.

 
Recalibration 

The QuadRTD or OctRTD standard calibration is at two points, 50 

 and 150 

.

  
Additional Services:

Custom calibration and verification point options available, please call for pricing.

 
Call for custom calibration options to accommodate specific application needs.
Prices and specifications subject to change. See MadgeTech’s terms and conditions at 

www.madgetech.com

To send devices to MadgeTech for calibration, service or repair, please use the MadgeTech RMA Process by visiting  

www.madgetech.com

, then under the services tab, select RMA Process.

3-Red

4-Red or Short to 3

1-Black or Short to 2

2-Black

Connector for 100
Ohm PtRTD

Connector for interface cable

3

4

1

2

3

4

1

2

KEY

4- Excitation Current Out (+)

3- Measurement (+) Input

2- Measurement(-) Input

1- Ground (-)