Device Operation (cont’d)
Trigger Settings
The device can be programmed to only record based off
user configured trigger settings.
1. In the
Connected devices
panel, select the intended
device to change the settings.
2. On the
Device
tab, in the
Information
group, click
Properties
. Users can also right-click on the device
and select
Properties
in the context menu.
3. Click
Trigger
and configure the Trigger settings.
Trigger formats are available in Window and Two Point
(bi-level) mode. Window mode allows for one range of
temperature monitoring and two point mode allows
for two ranges.
Set Password
To password protect the device so that others cannot
start, stop or reset the device:
1. In the
Connected Devices
panel, click the device
desired.
2. On the
Device
Tab, in the
Information
Group, click
Properties
. Or, right-click the device and select
Properties
in the context menu.
3. On the
General
Tab, click
Set Password
.
4. Enter and confirm the password in the box that
appears, then select
OK
.
Device Maintenance
O-Rings
O-ring maintenance is a key factor when properly
caring for the HiTemp140-FP. The O-rings ensure a tight
seal and prevent liquid from entering the inside of the
device. Please refer to the application note
O-Rings 101:
Protecting Your Data
, found at
, for
information on how to prevent O-ring failure.
Battery Replacement
Materials:
ER14250MR-145 Battery
1. Unscrew the bottom of the
logger and remove the battery.
2. Place the new battery into the
logger. Note the polarity of
the battery. It is important to
insert the battery with positive
polarity pointing upward
towards the probe. Failure to
do so could result in product inoperability or potential
explosion if exposed to high temperatures.
3. Screw the cover back onto the logger.
Recalibration
MadgeTech recommends annual recalibration. To send
devices back for calibration, visit
.
Notice: Steam Sterilization Applications
The pervasive nature of pressurized steam creates a very
difficult environment for electronics. Please refer to the
following preventative maintenance procedure when
using this device in steam sterilization applications.
Additionally, this device is not ideal for steam sterilization
applications above 121 °C/1.1 bar.
Preventive Maintenance
After every 3 hours of steam exposure:
1. Remove the endcap and battery from the device (ref.
battery change procedure on Product User Guide)
2. Place open logger (minus battery) in an oven at 120 °C
(250 °F) for a minimum of 30 minutes
3. Remove logger from oven and allow to cool to room
temperature
4. Re-assemble the logger with the battery (note polarity)
and endcap
Note:
This product is rated for use up to 140 °C
(284 ºF). Please heed the battery warning. The
product will explode if exposed to temperatures
above 140 °C (284 ºF).