Battery Replacement

Materials needed for the HiTemp140 and PR140: 

ER1425S-HT Lithium Battery

Battery Replacement Procedure: 

1. Unscrew the end cap of the data logger to expose the battery.
2. Remove the battery. 
3. Place the new battery into the data logger. Note the polarity of the battery. 
4. Tightly screw the data logger back together until the O-Rings are no longer visible.

BATTERY WARNING

: DISCARD USED BATTERY PROMPTLY. KEEP OUT OF REACH OF CHILDREN. DO NOT DISPOSE OF IN FIRE, RECHARGE, PUT IN 

BACKWARDS, DISASSEMBLE, OR MIX WITH OTHER BATTERY TYPES. MAY EXPLODE, FLAME, OR LEAK AND CAUSE PERSONAL INJURY.

O-Rings

O-Ring maintenance is a key factor when properly caring for the HiTemp140 and PR140. The O-Rings ensure a tight seal and prevent liquid from 
entering the inside of the device.

Materials needed for the HiTemp140:

 HiTemp140-O-Ring 

Materials needed for the PR140:

 PR140-O-Ring

O-Ring Replacement Procedure: 

1. Unscrew the end cap from the data logger to expose the O-Ring.
2. Use a small pointed tool (knife or pick) to pry the old O-Ring out of its groove.
3. Make sure that the O-Ring groove is free of any dirt or debris.
4. Partially screw the mating parts back together leaving the O-Ring groove exposed.
5. Apply a thin coat of silicone based lubricant to the O-Ring.
6. Stretch the O-Ring over the cap and into its groove. 

WARNING:

 Avoid stretching the O-Ring over the threads! Sharp threads can cut the new O-Ring.

7. Tightly screw the data logger back together until the O-Rings are no longer visible.

Maintenance: 

MadgeTech data loggers come directly from the factory with high quality O-Rings that have been properly installed. As a user, there are only a 
few things that need to be remembered to maintain a functional O-Ring seal.

DO:

• Clean them frequently (use compressed air or a soft brush to avoid abrasion).
• Lubricate regularly (if it doesn’t feel slippery, it needs to be lubricated). We recommend Parker® Super-O-Lube, but any silicone based O-Ring 
  lubricant will work. This is most important on the seals that are frequently opened and closed for communication with the logger.
• Inspect the O-Ring regularly for signs of failure.

DON’T: 

• Poke, jab, pry at the O-Ring with sharp or pointed objects.
• Expose the O-Rings to harsh chemicals.
• Expose the seals to high pressure (the HiTemp140 and PR140 data loggers are rated to 60 PSIG).
• Expose the seal to high temperatures (see specification for operating temperature range).

Data Logger Maintenance

Summary of Contents for AVS140-6

Page 1: ...AVS140 6 Autoclave Validation Data Logging System Product User Guide To view the full MadgeTech product line visit our website at madgetech com...

Page 2: ...on Protocol Software MIcro USB Cable 1 Included accessory with IFC406 Power Adapter 1 Included accessory with IFC406 Calibration Certificates 6 PR140 1 IFC406 1 HiTemp140 5 MadgeTech 4 Secure Software...

Page 3: ...direction making it easy to log temperatures within bottles vials or other hard to reach places HiTemp140 Product Overview HiTemp140 Specifications Hardware Temperature Sensor 100 Platinum RTD Probe M...

Page 4: ...0 0015 PSIA Calibrated Accuracy 0 03 Bar 0 435 PSI 20 C to 140 C 68 F to 284 F Response Time 0 1 ms 10 FSR to 90 FSR Reading Rate 1 second up to once every 24 hours Memory 65 536 readings Battery Lif...

Page 5: ...35 C 0 RH to 95 RH non condensing Connection Type USB to PC Weight 1 65 lb 750 g Material 6061 Aluminum PTFE impregnated hard anodized coating ABS Plastic Dimensions 9 5 in x 1 95 in x 1 75 in 241 3...

Page 6: ...integrity MadgeTech provides on site IQ OQ PQ services to help the user validate the data logging system Our trained staff are extremely knowledgeable with the standard IQ OQ PQ protocols of the Madge...

Page 7: ...nd numerous shortcuts while management will appreciate the increased data security and time saving elements Some of the most highly anticipated new features include Administrator User Settings Users c...

Page 8: ...that the data logger hardware is operational Performance Qualification PQ Recommendation Additional handling precautions for maintaining the accuracy of MadgeTech equipment Periodic maintenance inform...

Page 9: ...to the Windows Control Panel and select Programs Microsoft NET Framework 4 0 would be listed as a Windows program If the Windows PC doesn t already have NET 4 0 Framework click on MadgeTech Secure Sof...

Page 10: ...rt stores the dataset as it records while connected to the logger 2 The status of the device will change to Running Waiting to Start or Waiting to Manual Start depending upon your start method 3 Disco...

Page 11: ...f any dirt or debris 4 Partially screw the mating parts back together leaving the O Ring groove exposed 5 Apply a thin coat of silicone based lubricant to the O Ring 6 Stretch the O Ring over the cap...

Page 12: ...cure Software displays the date of the last calibration and the date that the device is next due for calibration The MadgeTech 4 Secure Software can also be configured to send an on screen notificatio...

Reviews: