PM100 Bump Test and Calibration Station
REV – 1.0
[34-2900-0513-2 ]
13
|
P a g e
5.3.1
Station Tab
The Station tab is displayed when the software opens (as shown above). It shows information about the Bump-
Calibration Station, information about the installed gas cylinder, configuration of the operation of the Bump-
Calibration Station, and the Bump-Calibration Station firmware upgrade interface.
NOTE
: No changes will be saved to the Bump-Calibration Station until the
Write
button is pressed (see 5.3.1.1).
5.3.1.1
Station Info
Serial Number
– Cannot be changed. It uniquely identifies the Bump-Calibration Station unit.
Version
– The firmware version installed on the Bump-Calibration Station.
Auto Power-off
– Determines the idle time before the Bump-Calibration Station automatically switches to sleep
mode. Can be set from 1-3600 seconds. Default is 300 seconds.
Maximum Bump Time
– Determines the maximum time allowed for a Bump Test. If the PM
R
100
R
device fails to
properly respond before the time expires, the test will stop, and the device will fail the test. Can be set from 20-120
seconds (default is 20 seconds).