PM400 Multi-Gas Personal Monitor
REV – 2.0
[34-2900-0206-2 ]
16
|
P a g e
Alarm
Alarm Standard
LCD Display
Alarms and Alerts
Calibration
Calibration due
Displays the specific
sensor(s) which
require calibration.
Perform a successful calibration to
clear.
Calibration
Failed
Failed calibration
Displays the specific
sensor(s) which
failed calibration.
Perform a successful calibration to
clear
4.7
Default Alarm Set Points
NOTE
: Alarm levels can only be changed using the PM Link and software, or the Bump Test-Calibration Station.
Gas Type
Combustible (LEL)
CO
H
2
S
O
2
Low Alarm
10% LEL
35 PPM
10 PPM
19.5%
High Alarm
20% LEL
100 PPM
20 PPM
23.5%
TWA Alarm
35 PPM
10 PPM
STEL Alarm
200 PPM
15 PPM
4.8
Bump Test Reminder
The Bump Test due reminder option can be enabled using the PM Link and software or the Bump Test-Calibration
Station. The Bump Test reminder can be configured from n/a (Off) to 365 days. The default configuration is set to
“Off”, no Bump Test reminder will be displayed.
When enabled, the monitor will prompt the user to perform the test by displaying “btS” on the display when the
test is due.