Manual
Instructions for use
DSPS & SSPS
14
02082021V19IFU112100
MacroMedics ®
9.
When using a mask on a pateint with long hair, place the opening in the frame around the
ponytail.
10.
Place the white frame slowly onto the black frame.
11.
Secure the mask in place by pressing down the plungers. Be careful not to pinch the patient’s
ears or skin around the neck.
12.
Use both hands to mould the structures of the face like the nose bone, orbitae, chin, and
forehead.
13.
If applicable, after moulding the facial region of the mask, proceed with moulding the mask
around the neck and shoulders.
14.
To speed up the setting process, you can use cold packs or cold, wet towels to cool the mask.
15.
Cure and cool the mask for at least 5 minutes.
16.
Release the plungers and remove the mask(s) carefully.
17.
Label the mask(s) with the patient’s name/ identifier.
9. Prone Treatment
Double Shell Positioning System (DSPS) for prone treatment