background image

DSPS

 

®

 

 Double Shell 

Positioning System 

&

SSPS

 

®

 

 Single Shell 

Positioning System

MacroMedics BV

Oostbaan 670

2841ML Moordrecht

The Netherlands

+31(0)182 389 777

[email protected]

www.macromedics.com

DSPS

SSPS

MANUAL 112100

Instructions for use

English Instructions

Summary of Contents for DSPS

Page 1: ... System SSPS Single Shell Positioning System MacroMedics BV Oostbaan 670 2841ML Moordrecht The Netherlands 31 0 182 389 777 info macromedics com www macromedics com DSPS SSPS MANUAL 112100 Instructions for use English Instructions ...

Page 2: ...Manual Instructions for use DSPS SSPS 2 02082021V19IFU112100 MacroMedics This page is intentionally left blank ...

Page 3: ...ts 1 Warnings 4 2 MRI Safety Information 4 3 Intended Use 5 4 Positioning Options 6 5 Cradle Set Up 8 6 Mask Set Up 11 7 Patient Set Up 11 8 Supine Treatment 12 9 Prone Treatment 14 10 Using shims 17 11 Cleaning Storage and Disposal 18 12 Warranty 18 13 Notices 18 14 Glossary of Symbols 19 ...

Page 4: ...t Do not remove the mask from the patient until the material cools completely Removing the thermoplastic mask before it has set may result in the mask s deformation If using more than one shim do not exceed a thickness of 1 5 mm In the USA federal law restricts this device to sale by or on the order of a licensed healthcare practitioner 2 MRI Safety Information 2 1 MR SAFE The following devices ar...

Page 5: ...dle Type O 111620 SSPS Cradle ExaFix 111630 SSPS Cradle U 112710 DSPS Cradle MEXF H 112720 DSPS Cradle O H 3 Intended Use Positioning of the patient during radiotherapy and radio diagnostics including MR where indicated 115510M Mask SSPS F 115520M Mask SSPS FS 115570M Mask SSPS FSL 115720M Mask SSPS FSS 115770M Mask SSPS FSSL 115810M Mask SSPS FVR 115820M Mask SSPS FSVRS 115860M Mask SSPS FVRL 115...

Page 6: ... Cradle Type ITV 112710 DSPS Cradle MEXF H 112720 DSPS Cradle O H 112730 DSPS Cradle S H 112740 DSPS Cradle VRT H 112750 DSPS Cradle B H 112760 DSPS Cradle I H 112770 DSPS Cradle U H Supine and Prone DSPS Cradle Positioning Options SSPS Cradle 111120 SSPS Cradle RE 111220 SSPS Cradle RE MR 111520 SSPS Cradle S Type 111530 SSPS Cradle S Type MR 111600 SSPS Cradle Type O Supine and Prone DSPS and SS...

Page 7: ...sk SSPS FSL 115620M Mask SSPS FSO 115670M Mask SSPS FSOL 114500M Mask SSPS FN 114550M Mask DSPS OS 114530M Mask UMCN Note MacroMedics offers thermoplastic masks with special cut outs for patients with long hair and for treatments using surface guided radiotherapy systems DSPS ONLY DSPS and SSPS 115720M Mask SSPS FSS 115770M Mask SSPS FSSL 115810M Mask SSPS FVR 115820M Mask SSPS FSVRS 115860M Mask ...

Page 8: ...QUICKFASTENERS I Ensure that the Plungers or Quickfasteners are in the released position II Align and slot the cradle onto the baseplate III Lock the cradle in place by pressing down on the plungers or quickfasteners IV After use the cradle can be detached by pulling on the plungers or quickfasteners and returning them to the released position Cradle Plunger Cradle Quickfastener in Released Positi...

Page 9: ...slot the cradle onto the baseplate III Lock the cradle in place by rotating the locking knobs 90 degrees so the knobs are pointing outwards IV After use the cradle can be detached by returning the locking knobs to the released position 5 3 TYPE B CRADLES I Attach the two pivots to the baseplate by screwing them into place Align and slot the cradle onto the baseplate ...

Page 10: ...ion to be able to place the reference box over the cradle Reference Box III When using the reference cradle place the reference box over the cradle as shown below Secure it in place by rotating the locking knobs 90 degrees so the knob is pointing towards the cranial end of the board IV After use the cradle and reference box can be detached by returning the locking knob to the released position and...

Page 11: ...ngers into the Plunger holders holes in the white frame of the mask Gently push them through until they are in released position 7 Patient Set Up Follow your department s patient positioning protocol for head and neck patients For DSPS treatments there should be approximately 1 finger width in between the cradle rim and the top on the patient s head For SSPS treatments position the patient by lett...

Page 12: ...e of the mask material 5 Lift the patient s head or ask the patient to sit up 6 Place the black frame onto the DSPS cradle with the MacroMedics logo facing upwards 7 Ensure that all of the frame s plungers are pushed properly into the holes in the cradle rim 8 Place the patient s head on the occipital mask and gently position the patient s head on top of the mask in the desired position for treatm...

Page 13: ...IAL MASK DSPS AND SSPS 1 Lay the patient in the supine position with the head resting on the occipital mask for DSPS or supine head support for SSPS 2 Take a new supine facial mask Perforated RayPRO thermoplastic mask with the white frame 3 Make sure that all 5 black plungers are in the released position 4 Heat the mask in a waterbath with a temperature between 68 C and 70 C 154 4 F and 158 F for ...

Page 14: ... 12 Use both hands to mould the structures of the face like the nose bone orbitae chin and forehead 13 If applicable after moulding the facial region of the mask proceed with moulding the mask around the neck and shoulders 14 To speed up the setting process you can use cold packs or cold wet towels to cool the mask 15 Cure and cool the mask for at least 5 minutes 16 Release the plungers and remove...

Page 15: ...achieve the desired height 4 Once the patient s head is at the correct height remove the Reference Bridge 5 Take a new prone facial mask Non perforated MacroCast thermoplastic mask with the black frame 6 Heat the mask in a waterbath with a temperature between 68 C and 70 C 154 4 F and 158 F for 2 to 4 minutes until it becomes sufficiently pliable 7 Take the mask out of the waterbath and dry it bet...

Page 16: ...NG A PRONE OCCIPITAL MASK DSPS AND SSPS 1 Lay the patient in the supine position with the head resting on the occipital mask for DSPS or prone head support for SSPS 2 Take a new prone occipital mask Perforated RayPRO thermoplastic mask with the white frame 3 Make sure that all 5 black plungers are in the released position 4 Heat the mask in a waterbath with a temperature between 68 C and 70 C 154 ...

Page 17: ...fort in the case of facial swelling the separation between the lower and upper masks may be increased by placing one or two shims between the masks Shims are available in 1 0 mm and 0 5 mm thickness The shims can be used for both supine and prone position An example of a shim used as part of the DSPS supine set up can be seen below Check the tab of the shim the tab is circled in blue below to iden...

Page 18: ...plicable to your country 12 Warranty The MacroMedics products are warranted to be free from defects in manufacturing material and construction The warranty is for 1 year starting from the date of delivery Product warranties remain valid provided the product was installed and or used according to the instructions Defects malfunctions or failures caused by nature misuse abuse and unauthorised altera...

Page 19: ...t cannot for a variety of reasons be presented on the medical device itself Prescription Only Devices for sale to physicians or trained healthcare providers only Keep Dry Indicates a medical device that needs to be protected from moisture Keep Away from Sunlight Indicates a medical device that needs protection from light sources GS1 Code 17 Use by Date Indicates the date after which the medical de...

Page 20: ...MacroMedics BV Oostbaan 670 2841ML Moordrecht The Netherlands 31 0 182 389 777 info macromedics com www macromedics com Visit www macromedics com for a digital copy of this manual ...

Reviews: