993065
–
Revision A
P a g e
|
8 of 12
Environmental - Storage and Operating Conditions
The Mackworth SA180 is intended for internal use within normal environmental conditions.
It is not intended to be used in environments where there are rapid changes in the environmental temperature and
humidity during intended use.
Normal Operating Conditions
+5°C to +40°C (41°F to 104°F) at a relative humidity between 15% to 90% RH, non-condensing but not requiring a water vapour
pressure greater than 50hPa and atmospheric pressure between 700hPa to 1060hPa.
Shipping and Storage Conditions
-25°C to +5°C (-13ºF to 41ºF) with any humidity level.
+5°C to +35°C (41°F to 95°F) at a relative humidity up to 90%.
+35°C to 70°C
(95⁰F to 158⁰F)
non-condensing at a water vapour pressure up to 50hPa.
Maintenance
No service is to be carried out on the Mackworth SA180 while transferring a person to reduce the risk of injury.
Service must be completed by a Prism Medical UK authorised Service Engineer.
Do not attempt to service the product yourself, or warranty is void.
To ensure the safety and continued good function of your Stand Aid, routine service must be performed on your Mackworth
SA180 Stand Aid.
Service should be completed by a Prism Medical UK approved service engineer every 6 months to ensure the products required
standard is maintained. The service history of the product should be documented each service in the Service Log at the back of
this User Manual.
When the Stand Aid is serviced, the 6 month service checklist must be completed for the Mackworth SA180 Stand Aid.
Service Manual Document Number: 995065
Spare Parts Manual Document Number: 992065
The Service must be completed every 6 months after installation of the Stand Aid to comply with LOLER Regulations.
The Mackworth SA180 Stand Aid has an expected Service Life of 10 Years.
Contact your local authorised Prism Medical UK dealer if you:
•
Need more information.
•
Have any questions about the use or service of your Stand Aid.
•
Notice any change in the performance.
•
Want to report an unexpected occurrence.
•
Want to arrange a service.
•
Need to ascertain necessary information for replacement parts and components.
Contact details of your local Prism Medical UK dealer are shown on the last page of this manual.