background image

Product Identification

Cautions

Product Cautions

Please use general safety precautions when
operating your Compressor and Nebulizer.

This unit should be used only for (1) its
intended purpose as described in this
guidebook, and (2) with medications only
under the supervision and instruction of your
physician.

READ THE FOLLOWING BEFORE USING

• To avoid electrical shock: Keep unit away

from water • Do not handle the unit or
power cord with wet hands • Do not
immerse the unit in liquid • Do not use
while bathing . • Do not reach for a unit that
has fallen into water – immediately unplug
the unit.

• Do not use the unit if it has any damaged

parts (including power cord or plug), if it
has been submersed in water or dropped.
Promptly send the unit for examination.

• The unit should not be used where

flammable gas, oxygen or aerosol spray
products are being used.

• Keep the air vents open. Do not place the

unit on a soft surface where the openings
can be blocked.

• If the medicine cup is empty, do not

attempt to operate the unit.

• If any abnormality occurs, discontinue use

until the unit has been examined and
repaired.

• The unit should not be left unattended

while plugged in.

• Do not tilt or shake the unit when in

operation.

• Disconnect the unit from the electrical

outlet before cleaning, filling and after
each use.

• Do not use attachments unless

recommended by the manufacturer.

Operating Cautions

• Close adult supervision is highly

recommended when this unit is used by
children or invalids.

• Keep your eyes away from the output of

medication mist.

• The maximum capacity of the medicine

cup is 13cc and should not be overfilled.

• Do not use this unit while operating a

vehicle.

• If any discomfort or abnormality occurs,

stop using the unit immediately.

Storage Cautions

• Do not store the unit in direct sunlight,

high temperature or humidity.

• Keep the unit out of reach of small

children.

• Always keep the unit unplugged while not

in use.

Cleaning Cautions

• Check the air filter, Nebulizer, mouthpiece,

and any other optional components before
each use. Dirty or worn parts should be
replaced.

• Do not immerse the compressor in water.

It may damage the unit.

• Disconnect the unit from the electrical

outlet before cleaning.

• Clean all necessary parts after each use

as instructed in this guidebook.

Cleaning Procedures

Please consult with your physician or
therapist prior to cleaning regarding the use
of commercial disinfectants and appropriate
cleaning methods.

3

Clear Cover

Filter Cover

Rechargable
Battery Pack

Nebulizer

POWER
Switch

Nebulizer

Battery
Terminal

AC port

Air Filter

Compressor

Rubber
Stem

Indicator
Light

Rubber Base

Summary of Contents for CompXP 40-136-000

Page 1: ...spiratory ailments in which an aerosolized medication is required during therapy Please consult with your physician and or pharmacist to determine if your prescription medication is approved for use w...

Page 2: ...e make no other express or implied warranties arising by operation of law or otherwise or any warranty of merchantability or fitness for a particular use or purpose whether or not the use or purpose h...

Page 3: ...tion Disconnect the unit from the electrical outlet before cleaning filling and after each use Do not use attachments unless recommended by the manufacturer Operating Cautions Close adult supervision...

Page 4: ...washer safe 3 Prepare a vinegar distilled water solution of one part white vinegar to three parts distilled water 4 Submerge the rinsed parts in the vinegar distilled water solution for 30 minutes 5 F...

Page 5: ...rely 5 Place the compressor on a solid flat dry surface Insert the nebulizer into the unit making sure that the stem of the medicine cup inserts securely into the rubber stem 2 Following general safet...

Page 6: ...therapist s instructions 5 Once the treatment is complete press the POWER switch to off and unplug the power cord from the outlet Remember Keep your eyes away from the output of medication mist If an...

Page 7: ...the outlet POWER switch is pressed No misting Check that there is or low rate medication in the Nebulizer of misting cup Check the Nebulizer for damage Check the position of the purple jet inside the...

Page 8: ...3 4 x 4 1 4 x 2 1 2 with cover and battery pack Approximate Weight 12 3 oz without battery pack 1 lbs 7 6 oz with battery pack Standard Accessories Portable handheld compressor clear cover nebulizer...

Page 9: ...con su m dico o con su farmac utico para determinar si la medicaci n que le han recetado est aprobada para ser utilizada con este nebulizador Este nebulizador est destinado al uso de un solo paciente...

Page 10: ...ientras se traslada Se recomienda un seguro de empaque con reci bo de devoluci n Si MABIS Healthcare determina que la unidad tuvo una falla en su funcionamiento debido a un defecto de fabricaci n la u...

Page 11: ...ntente operar la unidad Si ocurre cualquier hecho anormal inter rumpa el uso hasta que la unidad haya sido revisada y reparada No se deber a dejar la unidad sola mientras est enchufada No incline ni s...

Page 12: ...uave Enjuague con agua caliente del grifo NOTA Cuando limpie por primera vez o despu s de haber tenido la unidad guarda da por un per odo prolongado limpie con cuidado todos los componentes El nebuliz...

Page 13: ...que que el compresor y la bater a est n bien conectados NOTA Oprima el bot n de encendido Se iluminar una luz verde encima del bot n de encendido para indicar que la bater a est cargada La luz roja in...

Page 14: ...ci n de apagado Enchufe el cable de alimentaci n en un tomacorriente de la red 3 Coloque la llave interruptora en la posici n de encendido 4 Comience su tratamiento de acuerdo con las instrucciones de...

Page 15: ...la llave interruptora Falta de niebla Verifique que haya o flujo de niebla medicamento en el reducido vaso recipiente del nebulizador Verifique que el nebulizador no est da ado Verifique la posici n d...

Page 16: ...a 6 3 4 x 4 1 4 x 2 1 2 con tapa y bater a Peso aproximado 12 3 oz sin bater a 1 lb 7 6 oz con bater a Accesorios Compresor manual port til tapa transparente nebulizador manguera de 7 conector de man...