background image

Setup

General Guidelines

 

● Do not use the SNAP i.d.® 2.0 IHC system for 

deparaffinization, rehydration, dehydration, or mounting 

steps.

 

● Proper slide preparation is critical to successful use. 

Before

 placing tissue on slide, make sure that it will not 

extend beyond the target area of the assembly fixture.

 

IMPORTANT! 

Do not touch the gasket on the slide holder 

during assembly.

Slide Holder Setup

1.  Place assembly fixture on a hard, flat surface.
2.  Position slide in assembly fixture, with the tissue 

section facing up. Verify that the tissue section does 

not extend beyond the target area.

3.  Hold the slide holder by the sides, clip side down, and 

align it with the assembly fixture containing the slide. 

Do not touch the gasket

.

4.  Press the slide holder down firmly over the slide until 

the four compression clips slip over the slide edges 

with an audible click.

5.  To ensure that all clips are fully engaged, firmly press 

down on the top edge of each slide holder corner, one 

at a time, and listen for a click on each press. 

Click 4

Click 1

Click 2

Click 3

6.  Remove the slide holder assembly from the fixture 

and closely inspect each clip to confirm that it is 

in the proper position over the slide. If any clip is 

not engaging the slide as shown below, place the 

assembly back into the fixture and repeat step 5 on the 

unengaged clip(s).

Clip engaged

Clip not engaged

IHC Frame Setup

The IHC frame can be set up either on or off the 

SNAP i.d.® 2.0 system.

1.  Place assembled slide holders into any vacuum ports 

on the frame, with slide labels oriented towards the 

port numbers.

2.  Press blue port plugs 

lightly

 into all unused ports 

(excessive force is not required). Unused plugs can be 

left in the port plug storage holes.

3.  If the frame has been loaded off the SNAP i.d.

®

 2.0 

system, place it in either position on the system before 

continuing.

Staining Protocol

Tips for adding solutions to slide holder

 

● Small volumes (up to 600 µL) should be dispensed with 

a 1 mL pipette through the injection/recovery port. 

Gently

 push the pipette tip into the port to create a seal. 

Dispense slowly to avoid bubble formation in the tissue 

section chamber. 

 

● Larger volumes (up to 6 mL) can be loaded into the wash 

reservoir.

 

● Make sure that the fluid completely covers the tissue 

sections.

00119264_RevD.indd   2

1/14/2015   1:26:52 PM

Reviews: