Installation
Use the Installation Checklist and the Installation
Qualification Checklist (in this manual) as guides for
installing the mixer and and initiating use. Use the
Maintenance Checklist as a guide for followup care of
the mixer.
Install the GMP Mixer components in the following
order:
1. Tank plate (refer to the NovAseptic
®
Tank Plate
Welding Guide for installation instructions.)
2. Male bearing
3. Mixing head
4. Speed counter (optional, installed on the drive
unit)
5. Drive unit
Male Bearing Installation
See NovAseptic
®
Mixer GMP Male Bearing Installation Guide 00101868PU for additional
information.
• The vessel must be clean and completely dry inside.
• Do not use grip tools on bearing surfaces.
• Follow all local safety codes before entering the vessel.
• Handle the male bearing with care.
Mounting the Male Bearing
1. Ensure proper position of the O-ring on the male bearing.
2. For minimal friction, lubricate the visible part of the O-ring with a small amount of
purified water.
H
2
O
360°
!
2
3
5
1
4