SIMPLE LOOP TEST OUTPUT 

Option /E2

Simulated 0% and 100% output is available with Option /E2.  

 

Switch the Loop Test Selector positions.
  Top position:     0% fixed
  Middle position: 100% fixed
   Bottom position:  Cancel the loop test mode 

 

(Back to the normal operating mode)

CHECKING

1)  Terminal wiring: Check that all cables are correctly con

-

nected according to the connection diagram.

2)  Power input voltage: Check voltage across the terminal 

7 – 8 with a multimeter.

3)  Input:  Check  voltage  across  the  terminal  3  –  4  with  a 

sensitive voltmeter.

  (With 20°C or 68°F, approx. 220mV with Pt 100, approx.

  

110mV with Pt 50 Ω.)

  If RTD wires are broken, the output goes over 100% (be

-

low 0% with downscale protection) due to burnout pro

-

tection function. Check leadwires in such a case.

4)  Output:  Check  that  the  load  resistance  meets  the  de

-

scribed specifications.

ADJUSTMENT PROCEDURE

This unit is calibrated at the factory to meet the ordered 
specifications, therefore you usually do not need any cali-
bration.

The output signal can be finely adjusted to match it to a re

-

ceiving instrument, or to compensate input wire resistance 

when the unit is combined with a zenor barrier.  Follow the 
regular calibration procedure explained below:

 HOW TO CALIBRATE THE OUTPUT SIGNAL

Use a signal source and measuring instruments of sufficient 

accuracy level.  Turn the power supply on and warm up for 

more than 10 minutes.
1)  ZERO: Apply 0% input and adjust output to 0%.
2)  SPAN: Apply 100% input and adjust output to 100%.
3)  Check ZERO adjustment again with 0% input.
4)   When  ZERO  value  is  changed,  repeat  the  above  proce

-

dure 1) – 3).

MAINTENANCE

Regular calibration procedure is explained below:

 CALIBRATION

Warm up the unit for at least 10 minutes.  Apply 0%, 25%, 
50%,  75%  and  100%  input  signal.    Check  that  the  output 

signal for the respective input signal remains within the 

range  of  accuracy  described  in  the  data  sheet.   When  the 

output is out of tolerance, recalibrate the unit according to 

the “ADJUSTMENT PROCEDURE” explained earlier.

LIGHTNING SURGE PROTECTION

In order to protect the unit from lightning surges entering 
through signal and power supply cables, use of appropriate 
lightning surge protectors are recommended.  Please con-
tact M-System.