TERMINAL CONNECTIONS

Connect the unit as in the diagram below or refer to the connection diagram on the side of the unit.

 

EXTERNAL DIMENSIONS  unit: mm (inch)

5

4

6

7

1

2

3

[4 (.16)]

143 (5.63)

DIN RAIL

35mm wide

8 (.31)

27 (1.06)

39 (1.54)

• When mounting, no extra space is needed between units.

98 (3.86)

80 (3.15)

111 (4.37)

2–M4 MTG HOLE

7–M4 SCREW

10 (.39)

3

(.12)

TERMINAL COVER (option /T)

 

CONNECTION DIAGRAM

+

OUTPUT

4

5

1

2

LOAD

SOURCE

k

l

CT

L

K

U(+)

V(–)

6

7

AUXILIARY POWER

WIRING INSTRUCTIONS

 

SCREW TERMINAL 

Torque: 1.2 N·m

CHECKING

1) Terminal wiring:  Check that all cables are correctly con-

nected according to the connection diagram.

2) Auxiliary power input voltage:  Check voltage across the 

terminal 6 – 7 with a multimeter.

3) Input:  Check that the input signal is within 0 – 100% of 

full-scale.

  Be sure to short the secondary side of CT or turn off the 

current at its primary side for preventing the CT from 

burning when the cables are to be disconnected from in-

put terminals.

4) Output:  Check that the load resistance meets the de-

scribed specifications.

ADJUSTMENT PROCEDURE

This unit is calibrated at the factory to meet the ordered 
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case 
of regular calibration, adjust the output as explained in the 
following.

 

HOW TO CALIBRATE THE OUTPUT SIGNAL

Use a signal source and measuring instruments of suffi-
cient accuracy level.  Turn the auxiliary power supply on 
and warm up for more than 10 minutes.

1) ZERO:  Apply 0% input and adjust output to 0%.
2) SPAN:  Apply 100% input and adjust output to 100%.
3) Check ZERO adjustment again with 0% input.
4) When ZERO value is changed, repeat the above proce-

dure 1) – 3).

MAINTENANCE

Regular calibration procedure is explained below:

 

CALIBRATION

Warm up the unit for at least 10 minutes.  Apply 0%, 25%, 
50%, 75% and 100% input signal.  Check that the output 
signal for the respective input signal remains within accu-
racy described in the data sheet.  When the output is out of 
tolerance, recalibrate the unit according to the “ADJUST-
MENT PROCEDURE” explained earlier.

LIGHTNING SURGE PROTECTION

M-System offers a series of lightning surge protector for 
protection against induced lightning surges.  Please contact 
M-System to choose appropriate models.