LC2 / LCS2

EM-1357

P. 3 / 3

CHECKING

1) Terminal wiring:  Check that all cables are correctly

connected according to the connection diagram.

2) Power input voltage:  Check voltage across terminals 7 –

8 with a multimeter.

3) Strain gauge:  Check the bridge resistance.  The maximum

current supplied to the strain gauge is 65mA.

4) Input:  When applying maximum load, check that voltage

across terminals 10 (+) – 11 (–) equals Exc. 

×

 Strain Gauge

Sensitivity.

5) Output:  Check that the load resistance meets the de-

scribed specifications.

ADJUSTMENT PROCEDURE

• LCD Meter

Indicates input in % (0 – 100%).

• Excitation Adj.

The excitation voltage is calibrated at the factory, therefore
will not usually require adjustment, except when used in
combination with a strain gauge of sensitivity other than
described on the specification label.
In that case, adjust to the value calculated by the equation
below:

Excitation [V] =  5 [V] 

×

 a [mV/V]

          b [mV/V]

where
a: Sensitivity described on the specification label
b: User's sensitivity value

Adjustable range is within 2V to 5V with the bridge resist-
ance 350

.

• Zero Adj. (Tare Adj.)

Max. 80% of input span can be adjusted.
Without any load, adjust the output to 0%.

• Span Adj. (Gain Adjustment)

Max. 100 –␣ 20% of input span can be adjusted.
With the full load, adjust the output to 100%.

MAINTENANCE

Regular calibration procedure is explained below:

■ 

CALIBRATION

Warm up the unit for at least 10 minutes.  Apply 0%, 25%,
50%, 75% and 100% input signals.  Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet.  When the output is out of
tolerance, recalibrate the unit according to the "ADJUST-
MENT PROCEDURE" explained above.

Span Adj.

Zero Adj.

Excitation Adj.

LCD Meter

M-SYSTEM WARRANTY

M-System warrants such new M-System product which it manufactures to be free from defects in materials and workmanship during the 36-month period following the date that such
product was originally purchased if such product has been used under normal operating conditions and properly maintained, M-System's sole liability, and purchaser's exclusive remedies,
under this warranty are, at M-System's option, the repair, replacement or refund of the purchase price of any M-System product which is defective under the terms of this warranty.  To
submit a claim under this warranty, the purchaser must return, at its expense, the defective M-System product to the below address together with a copy of its original sales invoice.
THIS IS THE ONLY WARRANTY APPLICABLE TO M-SYSTEM PRODUCT AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.  M-SYSTEM SHALL HAVE NO LIABILITY FOR CONSEQUENTIAL, INCIDENTAL OR
SPECIAL DAMAGES OF ANY KIND WHATSOEVER.
M-System Co., Ltd., 5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN, Phone: (06) 6659-8201, Fax: (06) 6659-8510, E-mail: [email protected]

Reviews: