background image

48NAR

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201  Fax: +81(6)6659-8510  E-mail: [email protected]

EM-9432  Rev.8  P. 8 / 8

 

ZERO ADJUSTMENT

1) Apply 0% input signal.
2) Hold down [M] for 3 seconds* and push [M] to choose [Z].
3) Press [SET], and the bargraph shows 0%.

 

SPAN ADJUSTMENT

1) Apply 100% input signal.
2) Hold down [M] for 3 seconds* and push [M] to choose [S].
3) Press [SET], and the bargraph shows 100%.

 

ALARM SETPOINT ADJUSTMENTS

• 48NAR-2:  H 

[L setpoint] to 100%

   

0% to [H setpoint]

   

No alarm trip is selectable

• 48NAR-4:  HH  [H setpoint] to 100%
   

[L setpoint] to [HH setpoint]

   

[LL setpoint] to [H setpoint]

   

LL  0% to [L setpoint]

   

No alarm trip is selectable

• How to Set LL (or L, H, HH) Setpoints

1) Hold down [M] for 3 seconds* and push [M] to choose 

[LL].

2) Press [UP] or [DOWN] buttons until the indicator comes 

to the desired position, and press [SET].  

  (Same procedure for L, H, HH setpoints)

• How to Cancel HH and H Alarms

1) Press [UP] or [DOWN] buttons until the indicator comes 

to the upper limit, and press [SET].  

  HH indicator must be set before H as the H setpoint can-

not go above the HH setpoint.

• How to Cancel LL and L Alarms

1) Press [UP] or [DOWN] buttons until the indicator comes 

to the lower limit, and press [SET].  

  LL indicator must be set before L as the L setpoint can-

not go below the LL setpoint.

 

RESET TO FACTORY SETTING

1) Hold down [M] for 3 seconds* and press [M] to choose [•].
2) Press [UP] and [DOWN] buttons at once until the [•] 

LED starts blinking faster.

3) Press [SET] and [DOWN] buttons at once.

*  The [M] button remain activated for 1 minute after it has 

been used to choose a indicator LED (programming mode).  
During this period, pressing the [M] button briefly is enough 
to choose another indicator.  After 1 minute, it turns to deac-
tivated state.

CALIBRATION PROCEDURE

This unit is calibrated at the factory to meet the ordered 
specifications, therefore you usually do not need any cali-
bration.
The reading can be finely adjusted to match it to a receiving 
instrument, or to compensate input wire resistance when 
the unit is combined with a zenor barrier.  Follow the regu-
lar calibration procedure explained below:

 

HOW TO CALIBRATE THE OUTPUT INDICATION

Use a signal source and measuring instruments of sufficient 
accuracy level.  Turn the power supply on and warm up for 
more than 10 minutes.

1) ZERO:  Apply 0% input and adjust bargraph to 0% follow-

ing the procedure explained in “ADJUSTMENT PROCE-

DURE.”

2) SPAN:  Apply 100% input and adjust bargraph to 100% 

following the same procedure.

3) Check ZERO adjustment again with 0% input.
4) When ZERO value is changed, repeat the above proce-

dure 1) – 3).

MAINTENANCE

Regular checking procedure is explained below:

 

CHECKING

Warm up the unit for at least 10 minutes.  Apply 0%, 25%, 
50%, 75% and 100% input signal.  Check that the bargraph 
for the respective input signal remains within accuracy de-
scribed in the data sheet.  When the bargraph is out of toler-
ance, recalibrate the unit according to the “CALIBRATION 
PROCEDURE” explained earlier.

LIGHTNING SURGE PROTECTION

M-System offers a series of lightning surge protectors for 
protection against induced lightning surges.  Please contact 
M-System to choose appropriate models.

Summary of Contents for 48NAR

Page 1: ...figurations connected devices connected wires may affect the protection level of this unit when it is integrated in a panel system The user may have to review the CE or UKCA requirements in regard to the whole system and employ additional protective measures to ensure the CE or UKCA conformity In order to enable the operator to turn off the power in put immediately install a switch or a circuit br...

Page 2: ...NENT IDENTIFICATION 50 40 30 20 10 0 10 20 30 40 50 m3 h 100 SET H S Z L 90 80 70 60 50 40 30 20 10 0 M Brackets Body A Scale Plate Pull up the part A when replacing the scale plate INSTALLATION PANEL CUTOUT unit mm 138 1 0 164 min Note 1 Observe at the minimum of 3 cm above and below the units for heat dissipation Note 2 No bezel is needed when the panel is cut according to the above drawings Pan...

Page 3: ... mm Slide the top and bottom parts of the bezel into the back of the front cover as shown below Bezel When multiple units are to be clustered side by side insert one bezel between two units as shown below When mounting two or more indicators side by side HOW TO MOUNT THE UNIT ON A PANEL 1 Insert the unit into the panel cutout The watertight packing must be in place to hold the unit Do not remove i...

Page 4: ... 13 14 15 16 16 M3 SCREW TERMINALS Rounded corners for the option D 2 M3 SCREWS 144 5 67 137 5 5 41 4 1 16 36 1 42 OPTION D BEZEL 3 1 Space required when replacing the scale plate 2 Used for the existing panel cutout of M System 48 Series 38 139 5 mm 3 Used for the existing DIN panel cutout 33 138 mm 4 Used for the existing panel cutout of Fuji Electric PAJ PAK PBA 44 138 mm etc STANDARD BEZEL 2 1...

Page 5: ...owered 7 8 9 8 9 10 HHa 7 Hc HHc HHb Ha HH OUTPUT H OUTPUT LL OUTPUT L OUTPUT ALARM OUTPUT CODE 2 2 points ALARM OUTPUT CODE 4 4 points A B B RTD A B B RTD U V La Lb Lc Hc Ha Hb H OUTPUT L OUTPUT POWER U V POWER 4 3 2 4 3 2 WIRING INSTRUCTIONS SCREW TERMINAL Torque 0 6 N m SOLDERLESS TERMINAL Refer to the drawing below for recommended ring tongue terminal size Spade tongue type is also applicable ...

Page 6: ...cator amber Zero Adj amber L Setpoint Adj Trip Indicator green UP Button Mode Selector DOWN Button H Setpoint L Setpoint PV Indicator M 50 40 30 20 10 0 10 20 30 40 50 C 100 SET L Z S 90 80 70 60 50 40 30 20 10 0 H M Vertical Mounting Horizontal Mounting 50 40 30 20 10 0 10 20 30 40 50 C 100 SET H S Z L 90 80 70 60 50 40 30 20 10 0 SET ENTER Button HH Setpoint Adj Trip Indicator red H Setpoint Adj...

Page 7: ...djustment ALARM SETTING Proceed after the zero span adjustments 1 Keep pressing at least for 3 seconds to activate Mode Selector M Press briefly for second and more times within 1 minute after it has been activated 2 HH or LL setpoints are not provided for the 48NAR 2 3 The reset function is used to reset the unit to the factory set temperature range It does not affect alarm setting Note Each sett...

Page 8: ...at once The M button remain activated for 1 minute after it has been used to choose a indicator LED programming mode During this period pressing the M button briefly is enough to choose another indicator After 1 minute it turns to deac tivated state CALIBRATION PROCEDURE This unit is calibrated at the factory to meet the ordered specifications therefore you usually do not need any cali bration The...

Reviews: