background image

MODEL: 47NLN

47NLN SPECIFICATIONS

ES-9563 Rev.4   Page 1/4

https://www.m-system.co.jp/

Digital Panel Meters 47NL Series

LOOP POWERED DIGITAL PANEL METER

(4 1/2 digit, LED display type)
Functions & Features
• 4 ½ digit loop powered digital panel meter
• No external power source needed
• Moving average function to suppress the display flickering
• Scaling, forced zero, low-end cutout
• Max. and Min. value display

Zro

Spn

D/P

Tch

Fnc

Min

Max

Scale/

Max/Min

Shift

Up

96.0 (3.78)

48.0

(1.89)

20.0 (.79)

mm (inch)

MODEL: 47NLN-[1][2]

ORDERING INFORMATION

• Code number: 47NLN-[1][2]
Specify a code from below for each of [1] and [2].
  (e.g. 47NLN-R/Q)
• Specify the specification for option code /Q
  (e.g. /SET)

INPUT

Current

4 – 20 mA DC

[1] DISPLAY COLOR

R: Red
G: Green

[2] OPTIONS

blank: none
/Q: With options (specify the specification)

SPECIFICATIONS OF OPTION: Q

EX-FACTORY SETTING

/SET: Preset according to the Ordering Information Sheet
 (No. ESU-9563)

GENERAL SPECIFICATIONS

Construction: Panel mount type
Degree of protection: IP66; Applicable to the front of the
panel meter mounted according to the specified panel
cutout. Only screw mounting conforms.
Connection: Screwless spring terminal
Applicable wire size: 0.2 to 0.5 mm

2

, stripped length 6 mm

Housing material: Flame-resistant resin (gray)
Setting: (Front button)
• Scaled range
• Moving average
• Others
(Refer to the instruction manual for details)
Averaging: None or moving average
Lockout setting: Prohibiting certain operations; protecting
settings

DISPLAY

Display: 16 mm (.63) high, 4 ½ digits, 7-segment LED
Display range: -19999 to 19999
Decimal point position: 10

–1

, 10

–2

, 10

–3

, 10

–4

 or none

Zero indication: Higher-digit zeros are suppressed.
Over-range indication: ‘-19999’ or ‘19999’ blinking for
display values out of the display range.
‘S.ERR’ and ‘Min’ or ‘Max’ blinking when the input signal is
out of the usable range.
Function indicators:
Zro, Spn, D/P, Tch, Fnc, Min, Max
Display mode status and operation status, ON or blink
(Display color is the same as display color code.)
Engineering unit indication: Sticker label attached
DC, AC, mV, V, kV, μA, mA, A, kA, mW, W,
kW, var, kvar, Mvar, VA, Hz, Ω, kΩ, MΩ,
cm, mm, m, m/sec, mm/min, cm/min, m/min,
m/h, m/s

2

, inch, ℓ, ℓ/s, ℓ/min, ℓ/h, m

3

, m

3

/sec,

m

3

/min, m

3

/h, Nm

3

/h, N·m, N/m

2

, g, kg, kg/h,

N, kN, Pa, kPa, MPa, t, t/h, ℃, °F, %RH, J,
kJ, MJ, rpm, sec, min, pH, %, ppm, etc.

INPUT SPECIFICATIONS

■ DC Current
Current range: Approx. 3.7 to 23 mA DC
Voltage drop: Approx. 6.3 V with 4 mA; approx. 6.5 V with
20 mA (Equivalent input impedance: Approx. 325 Ω)
(There is voltage drop generated by using the unit.  For the
two-wire transmitter power supply, make sure that the
operable voltage for the two-wire transmitter is ensured
including the voltage drop by other devices and wiring
resistance.)

Summary of Contents for 47NL Series

Page 1: ... resin gray Setting Front button Scaled range Moving average Others Refer to the instruction manual for details Averaging None or moving average Lockout setting Prohibiting certain operations protecting settings DISPLAY Display 16 mm 63 high 4 digits 7 segment LED Display range 19999 to 19999 Decimal point position 10 1 10 2 10 3 10 4 or none Zero indication Higher digit zeros are suppressed Over ...

Page 2: ...DENTIFICATION No COMPONENT FUNCTION 1 Main display Indicates present values setting values and status of the unit 2 Function indicators Indicates the status in each setting mode 3 Max Min button Used to switch the main display to show the present values maximum values or minimum values 4 Button Used to shift through setting items in each setting mode etc 5 Scale button Used to move on to the scali...

Page 3: ...protection IP66 Panel cutout must be such as M System specified Use watertight packing and unit fixing screws included in the product package 42 25 1 66 42 25 1 66 4 4 5 1 8 d i a 20 5 81 17 67 3 0 5 1 2 0 d i a Single Mounting unit mm inch MAGNET MOUNTING Not conform to degree of protection IP66 Panel cutout is not required Use magnet sheet and stopper included in the product package Clustered Mo...

Page 4: ...jp SCHEMATIC CIRCUITRY CONNECTION DIAGRAM 2 1 INPUT Two wire Transmitter Power Source Digital Computation A D Converter Display Setting 3 4 5 6 7 8 Controllor etc 4 20mA DC 2 1 INPUT 4 WIRE CONNECTION EXAMPLE I P Positioner etc Specifications are subject to change without notice ...

Reviews: