MAINTENANCE

Regular calibration procedure is explained below:

 CALIBRATION

Warm up the unit for at least 10 minutes.  Apply 0%, 25%, 
50%,  75%  and  100%  input  signal.    Check  that  the  output 

signal for the respective input signal remains within accu-
racy described in the data sheet.

M-SYSTEM WARRANTY

M-System warrants such new M-System product which it manufactures to be free from defects in materials and workmanship during the 36-month period following the date that such 
product was originally purchased if such product has been used under normal operating conditions and properly maintained, M-System’s sole liability, and purchaser’s exclusive 
remedies, under this warranty are, at M-System’s option, the repair, replacement or refund of the purchase price of any M-System product which is defective under the terms of this 
warranty.  To submit a claim under this warranty, the purchaser must return, at its expense, the defective M-System product to the below address together with a copy of its original 
sales invoice.
THIS IS THE ONLY WARRANTY APPLICABLE TO M-SYSTEM PRODUCT AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY IMPLIED 
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.  M-SYSTEM SHALL HAVE NO LIABILITY FOR CONSEQUENTIAL, INCIDENTAL OR 
SPECIAL DAMAGES OF ANY KIND WHATSOEVER.
M-System Co., Ltd., 5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN, Phone: (06) 6659-8201, Fax: (06) 6659-8510, E-mail: [email protected]

+

+

+

4

1

5

6

17

16

15

14

+15V

15V

+

 Non-inverting amplifier circuit: G = 0.95 (Isolation: Input to output or power)

INPUT ±5V DC

POWER 15V DC

OUTPUT approx. ±4.75V DC

+

+

+

16

17

14

15

1

4

6

5

+

 Non-inverting amplifier circuit: G = 0.95 (Isolation: Output to input or power)

INPUT ±5V DC

POWER 15V DC

OUTPUT approx. ±4.75V DC

A1

+

REFERENCE VOLTAGE SOURCE

Note: The amplifier A1 requires a power source independent from the Reference Voltage Source.

The primary amplifier in this unit is high accurate.