WIRING INSTRUCTIONS

 

SCREW TERMINAL

Torque: 0.8 N·m

CHECKING

1) Terminal wiring:  Check that all cables are correctly con-

nected according to the connection diagram.

2) Power input voltage:  Check voltage supplied to the rack 

(model: 10BXx).  For the DC power source, be sure that 

the ripple level is within 10% p-p.

3) Input:  Check that the input signal is within 0 – 100% of 

the full-scale.

4) Output:  Check that the load resistance meets the de-

scribed specifications shown below.

OUTPUT

LOAD REQUIREMENTS

Open collector

30V DC @ 100mA max.

5V pulse

250 Ω min.

Photo MOSFET 
relay pulse

30V DC @200mA max. (resistive load)
132V AC @200mA max. (cosø=1)

  Check also that the receiving instrument is able to accept 

the waveform as explained in the figure below.

• Open Collector

• Voltage Pulse

OFF

ON

20%

80%

1 / f

 Frequency less than 500 Hz at 100% input

        

 Duty ratio 20% (See the figure below)

• OUTPUT 2’s Open Collector and Photo MOSFET 
  Relay Pulse

OFF

ON

20%

80%

1 / f

H Level

L Level

20%

80%

1 / f

• Open Collector

• Voltage Pulse

 Frequency greater than 500 Hz at 100% input

        

 See the figure and equation below.

H Level

L Level

Pulse Width

OFF

ON

Pulse Width

Pulse Width [millisec.] =

1

2.09 × 100% Frequency [kHz]

 When OUTPUT 1 is Photo MOSFET Relay Pulse

        

 See the figure below.  ON pulse width is limited 

            within 75 ±25 msec. when the output frequency 
            gets low (below 2 to 4 Hz).

ADJUSTMENT PROCEDURE

This unit is calibrated at the factory to meet the ordered 
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case 
of regular calibration, adjust the output as explained in the 
following.

 

HOW TO CALIBRATE THE OUTPUT SIGNAL

Use a signal source and measuring instruments of sufficient 
accuracy level.  Turn the power supply on and warm up for 
more than 10 minutes.

1) ZERO:  Apply 5% input and adjust output to 5%.
2) SPAN:  Apply 100% input and adjust output to 100%.
3) Check ZERO adjustment again with 5% input.
4) When ZERO value is changed, repeat the above proce-

dure 1) – 3).

MAINTENANCE

Regular calibration procedure is explained below:

 

CALIBRATION

Warm up the unit for at least 10 minutes.  Apply 0%, 5%, 
25%, 50%, 75% and 100% input signal.  Check that the out-
put signal for the respective input signal remains within 
accuracy described in the data sheet.  When the output is 
out of tolerance, recalibrate the unit according to the “AD-
JUSTMENT PROCEDURE” explained earlier.

LIGHTNING SURGE PROTECTION

M-System offers a series of lightning surge protector for 
protection against induced lightning surges.  Please contact 
M-System to choose appropriate models.