Lynton Luminette Q Operator'S Manual Download Page 20

16 

________________________________________________________________________________ 

are  completely  pushed  in  (they  should  latch  in  position).  Then,  line  up  the 

electrical and water connectors and push the plug into the socket until you 
hear the two clicks of the water connector catches snapping into position. It 

may  be  necessary  to  hold  the  Luminette  Q  System  steady  with  your  other 
hand during this operation. 

 

3.6  Interlock Warning Screen and LEDs  

 
If  at  any  stage  a  fault  is  detected,  the  Luminette  Q  Laser  System  will 

automatically take appropriate action to render itself safe and a message will 

be displayed on the screen. If the External Interlock (i.e. the Door Interlock), 
or the System Temperature Interlock are broken it will be impossible to fire 

the Flashlamp and the screen will change to: 
 

INTERLOCK

 

BROKEN 

Door

-

Interlock 

                                         

Close the Door & Press

-

any

-

Key 

 

If,  however,  there  is  a  more  serious  Water  Flow,  Water  Level  or  Water 

Temperature  fault,  the  Flashlamp  Simmer  will  also  be  extinguished  and  the 
screen will change to: 

 

INTERLOCK

-

BROKEN 

Water

-

Level or Flow

-

Problem 

Check LED status at the back 

Turn

-

OFF

-

& Call Lynton on +441477536977

-

 

 
The internal temperature of the system is monitored and if found to be too 

high, the following screen is displayed:  

 

Or, more specifically, 
 

INTERLOCK

 

BROKEN 

Simmer

-

Problem 

                                         

Turn

-

OFF & Call Lynton on +441477536977

 -

 

 

WARNING : Failure to ensure that the water connectors have been correctly 

“snapped”  shut  could  result  in  the  Handpiece  becoming  detached  during 

use.  Interlocks  would  ensure  that  the  system  turns  itself  off  and  remains 

safe, but damage could be done to the connector itself. 

INTERLOCK

-

BROKEN 

High System Temperature 

 

Leave System to Cool & Press any Key

-

 

Summary of Contents for Luminette Q

Page 1: ......

Page 2: ......

Page 3: ...he Luminette Q Laser System 12 3 5 Changing Handpieces 15 3 6 Interlock Warning Screen and LEDs 16 4 CLEANING AND MAINTENANCE 19 4 1 Cleaning 19 4 2 Maintenance Schedule 19 4 3 Maintenance Procedures...

Page 4: ...A1 3 Available Options and Accessories 22 A1 4 Labelling 23...

Page 5: ...the Luminette range of IPL laser systems the standard Luminette System the Luminette Advance system and the recently introduced Luminette Q system All of these models look similar from a physical and...

Page 6: ...ttons Pre set default parameters are included for the different Skin Types all determined through clinical experience The output of the Luminette Q Laser System comprises of passively Q switched laser...

Page 7: ...onent Users are particularly recommended to read the accompanying Training Manual before using the Luminette Q System for its intended purposes in cosmetic dermatology CAUTION The Luminette Q Laser Sy...

Page 8: ...detecting equipment should be used The radiation classed as completely invisible may interact with any surface on which it impinges to produce visible light The user is warned that even this light may...

Page 9: ...chable Handpiece properly connected it does not present the operator with any electrical hazard When Detachable Handpieces are being changed over there are no hazardous voltages on any of the electric...

Page 10: ......

Page 11: ...ected to ensure that it has not been unreasonably damaged during transit If the crate has been damaged it is recommended that the supplier is contacted immediately It will then be recommended that pho...

Page 12: ...Section 4 3 2 of Chapter 4 Cleaning Maintenance Because the internal cooling system requires an air vent a small hole in the pipe of the reservoir filling point to operate properly it is likely that...

Page 13: ...ressed The key cannot be removed in this position Turn the key switch counter clockwise to turn the Luminette Q System off again at any time Remove the key and store in a safe location when the Lumine...

Page 14: ...e Q System is shown below 6 Fused Mains Inlet Pulling out the detachable mains lead from this socket removes all power from the Luminette Q System Alternatively the mains lead can unplugged from the w...

Page 15: ...ividual LEDs comes later 10 Reservoir Filling Point The internal cooling system can be topped up at this point using the pipe provided The pipe is normally fitted with a two position cap In normal use...

Page 16: ...ns performed at any time will remove the possibility of the emission of light i Turning the key switch counter clockwise to the OFF position ii Pressing the Emergency Stop Button iii Switching off or...

Page 17: ...ers can be selected and the system can be enabled for treatment This Standby screen displays the handpiece type currently connected to the Luminette Q system in the HP area of the screen As can be see...

Page 18: ...ing this button increases the rate When the maximum Rate is reached the setting then reverts back to the minimum Rate At this point the User can fire the handpiece by pressing the Pushbutton on the Ha...

Page 19: ...to remove a Detachable Handpiece and re connect another one To remove a Detachable Handpiece it is necessary to simultaneously press the two buttons on either side of the connector socket whilst at th...

Page 20: ...le to fire the Flashlamp and the screen will change to INTERLOCK BROKEN Door Interlock Close the Door Press any Key If however there is a more serious Water Flow Water Level or Water Temperature fault...

Page 21: ...hot Inhibits Flashing If the fault cannot be easily corrected contact Lynton Lasers Ltd or one of its designated agents stating which LED s is are lit If a handpiece reaches the end of it s recommende...

Page 22: ...18 ________________________________________________________________________________...

Page 23: ...rial e g Circuit Diagrams Mechanical Design Drawings etc can be made available after consultation with Lynton Lasers Ltd Service Contracts are also available 4 3 1 Topping up the Cooling Water If ther...

Page 24: ...will light up as well as No 1 Remove the black Reservoir Filler Cap from the Reservoir Filling Point on the back panel of the Luminette Q System and fill up with deionised water using the Filling Bott...

Page 25: ...able anaesthetic mixture with air or with oxygen or with nitrous oxide NO CLASSIFICATION Classification according to the mode of operation CONTINUOUS OPERATION The Equipment is also classified as PORT...

Page 26: ...r a suitable pair of glasses or goggles for the Luminette Q Laser System contact Lynton Lasers Ltd A1 2 5 Electrical Parameters Operating Voltage Range 100 125 or 200 264 Volts AC 2 engineer selectabl...

Page 27: ...ating Plate panel of the label B Panel B Rating Plate panel containing the rated mains voltage with the sign indicating that it is an AC voltage the input mains power the input mains frequency the Ele...

Page 28: ...t the Luminette Q is a Class 4 Laser System Laser Radiation Starburst Hazard Label Label indicating the requirement for efficient recovery of the Luminette Q System at end of use according to the WEEE...

Page 29: ...Lynton House Manor Lane Holmes Chapel Cheshire CW4 8AF United Kingdom Telephone 44 0 1477 536977 Fax 44 0 1477 536978 email Info lynton co uk...

Reviews: