20 AIM-100™ Operation and Maintenance Manual – 1008189 A03
IEC60320C13, length: 2,5 m.
Transport and Storage Conditions
Ambient temperature
0 - 70
°
C (32 - 158
°
F)
Relative humidity
10 - 75% (no condensation)
Atmospheric pressure
500 - 1060 hPa
Operation Conditions
Ambient temperature
10 - 30
°
C (50 - 86
°
F)
Relative humidity
30 - 75%
Atmospheric pressure
700 -1060 hPa
Weights and Dimensions
Single ceiling version
33.6 lbs, (15,3 kg)
Double ceiling version
70 lbs (32 kg)
Wall mount version
40 lbs (18 kg)
Floor stand version
34.8 lbs (15,8 kg)
Light head including yoke
9,6 lbs (4,4 kg)
Light head diameter
20 in (51 cm)
Range of motion:
See illustrations in Technical Description.
Optical Performance/Operating Characteristics
The optical data are nominal values (unless specified otherwise) based on
measurements done according to the IEC 60601-2-41 standard on the rated voltage.
Please be aware that lights are very sensitive to voltage variations, and that this can
affect the light performance and bulb life significantly. Also the performance of the
individual bulbs has a natural variance due to their manufacturing process.
Central Illumination at 1 m (39.4”) , E
c
65,000 lux (6000 fc)
Central Illumination at 24” (0.61 m) , E
c
160,000 lux (14800 fc)
Light Field Diameter, d
10
9.7 in (246 mm)
5
Light Field Diameter, d
20
7.3 in (185 mm)
3
Light Field Diameter, d
50
4.5 in (114 mm)
3
Light distribution ratio d
50
/d
10
0.5
Total Irradiance, E
e
380.0 W/m
2
(0.04 W/cm
2
)
Depth of illumination
47 in (1,2 m)
Color temperature
3700°K
CRI (Color Rendering Index)
97 R
a
R9 (Color Rendering of a strong red color that is
important for medical purposes)
95 R
9
Focus range
24in (0,61m) to 39in (1m)
Remaining Illuminance With Two Masks
21 % of Ec
Remaining Illuminance With One Tube And Two Masks 32 % of Ec
Remaining Illuminance With One Mask
16 % of Ec
Remaining Illuminance With Tube (Deep Cavity)
93 % of Ec
Remaining Illuminance With One Tube And One Mask
21 % of Ec
Cavity Shadow Ratio
0.21 T
Ecological characteristics
Burton Medical Products cares about the ecological environment, and in the design
5
The Light Field Diameters are given for a light beam that is merged for maximum central illuminance.
Since the light beam is adjustable, the Light Field Diameter can be increased by offsetting the merging of
the beams. See more about this in the section “Adjusting the light pattern” on page 9.