"% Salt" compensation value adjustment
If intend to change the % Salt compensation value,
refer page 38, chapter 9-8 ( % Salt Compensation
value Setting ).
"Height" compensation value adjustment
If intend to change the Height compensation value,
refer page 39, chapter 9-9, 9-10 ( Height Compensation
value Setting ).
7-2 Calibration
1)Install the " Probe Plug " ( 6-1, Fig. 4 ) into the
" DO Socket " ( 3-13, Fig. 1 ).
2)Power on the meter by pressing " Power Button "
( 3-2, Fig. 1 ) once.
Select the measurement mode to " Dissolved Oxygen
measurement ", the Display will show " mg/L " indicator.
Pressing the " Function Button " ( 3-6, Fig. 1 ) once
until the right bottom display show " %O2 ".
3)Wait for approx. 5 minutes at least until the
display reading values become stable & no
fluctuation.
4)Use the two fingers to press the " REC Button " ( 3-4,
Fig 1 ) and HOLD Button " ( 3-3, Fig. 1 ) at the same
time, the display will show the following screen as
example, release the both fingers.
20.9
CAL
%O2
28