7-4 Set beeper sound ON/OFF

When the lower display show " bEEP " 

1)Use the "   Button " ( 3-6, Fig. 1 ) or  "   Button " 

( 3-5, Fig. 1 ) to select the upper value to " yES " or
" no ".

yES - Meter's beep sound will be ON with default.
no - Meter's beep sound will be OFF with default.
          is power ON.

2)After select the upper text to " yES " or " no ",  press the 

" Enter Button " ( 3-4, Fig. 1 ) will  save the setting 
function with default.

7-5 Select the Thermometer to Type K or Type J

When the lower display show " tYPE " 

1)Use the "   Button " ( 3-6, Fig. 1 ) or  "   Button " 

( 3-5, Fig. 1 ) to select the Display unit to " K " or
" J ".

K - Type K thermometer
j - Type J thermometer

2)After Display unit is selected to " K " or " J ",  press the 

" Enter Button " ( 3-4, Fig. 1 ) will  save the setting 
function with default.

19

Summary of Contents for UVC-254SD

Page 1: ...LIGHT METER with SD CARD DATALOGGER marks a step forward for you into the field of precision measurement Although this METER is a complex and delicate instrument its durable structure developed Pleas...

Page 2: ...4 2 Data Hold 9 4 3 Data Record Max Min reading 10 4 4 LCD Backlight ON OFF 10 5 DATALOGGER 11 5 1 Preparation before execute datalogger function 11 5 2 Auto Datalogger Set sampling time 1 second 12...

Page 3: ...ermocouple thermometer Real time SD memory card Datalogger it Built in Clock and Calendar real time data recorder sampling time set from 1 second to 3600 seconds Manual datalogger is available set the...

Page 4: ...SPECIFICATIONS 2 1 General Specifications Circuit Custom one chip of microprocessor LSI circuit Display LCD size 52 mm x 38 mm LCD with green backlight ON OFF Measurement UVC Light Type K J thermomete...

Page 5: ...of SD card setting Auto power OFF management Set beep Sound ON OFF Set thermometer type to Type K or Type J Set temperature unit to or Set sampling time SD memory card Format Temperature Automatic tem...

Page 6: ...Dimension Meter 177 x 68 x 45 mm UVC probe 38 mm DIA x 25 mm Accessories Instruction manual 1 PC Included UVC light sensor 1 PC Hard carrying case CA 06 1 PC Optional SD Card 2 GB Accessories Type K t...

Page 7: ...o 1300 0 0 2 0 5 50 1 to 100 0 0 2 1 0 1 58 0 to 2372 0 0 2 1 58 1 to 148 0 0 2 1 8 Type J 0 1 50 0 to 1100 0 0 2 0 5 50 1 to 100 0 0 2 1 0 1 58 0 to 2012 0 0 2 1 58 1 to 148 0 0 2 1 8 Above specifica...

Page 8: ...Button 3 6 Logger Button Button Sampling time check Button Zero Button 3 7 Probe input socket 3 8 Type K J Temp Probe Input Socket 3 9 SD card socket 3 10 RS 232 Output Terminal 3 11 Reset Button 3 12...

Page 9: ...f the meter 2 The meter can select 2 kind function as a UVC Light meter b Type K J Thermometer Pressing the Function Button 3 3 Fig 1 continuously not release the button the Display will show the foll...

Page 10: ...unit Now the meter is ready for making the UVC light measurement d Use the fingers to hold the UVC Sensor body 3 18 face the UVC Sensor 3 19 Fig 1 against to the measurement UVC light source directly...

Page 11: ...ing from the Temp probe d If the Display show the indicator K it is ready for Type K thermometer If the Display show the indicator J it is ready for Type J thermometer Remark The meter is defaulted to...

Page 12: ...e memory function continuously b Press the REC Button 3 4 Fig 1 again the REC MIN symbol along with the minimum value will appear on the display If intend to delete the minimum value just press the Ho...

Page 13: ...commend strongly do not use memory cards that have been formatted by other meter or by other installation such as camera Reformat the memory card with your meter c Time setting If the meter is used at...

Page 14: ...nable refer to Chapter 7 4 page 19 b Pause the datalogger During execute the Datalogger function if press the Logger Button 3 6 Fig 1 once will pause the Datalogger function stop to save the measuring...

Page 15: ...5 Fig 1 the lower no position no will flashing It can use the Button 3 6 Fig 1 or Button 3 5 Fig 1 to set the measuring position 1 to 99 for example room 1 to room 99 to identify the measurement loca...

Page 16: ...the first time to execute the Datalogger under the route UVC01 will generate a new file name UVC01001 XLS After exist the Datalogger then execute again the data will save to the UVC01001 XLS until Dat...

Page 17: ...SD card into the SD card adapter then connect the SD card adapter into the computer 3 Power ON the computer and run the EXCEL software Down load the saving data file for example the file name UVC01001...

Page 18: ...ess the Next Button 3 3 Fig 1 once a while in sequence to select the eight main function the lower display will show dAtE Set clock time Year Month Date Hour Minute Second dEC Set SD card Decimal char...

Page 19: ...st the value Setting start from Year value After the desired value is set press the Enter Button 3 4 Fig 1 once will going to next value adjustment for example first setting value is Year then next to...

Page 20: ...er value to bASIC or Euro bASIC Use as the Decimal point with default Euro Use as the Decimal point with default 2 After select the upper text to bASIC or Euro press the Enter Button 3 4 Fig 1 will sa...

Page 21: ...N 2 After select the upper text to yES or no press the Enter Button 3 4 Fig 1 will save the setting function with default 7 5 Select the Thermometer to Type K or Type J When the lower display show tYP...

Page 22: ...7 Set sampling time SecondS When the lower display show SP t 1 Use the Button 3 6 Fig 1 or Button 3 5 Fig 1 to adjust the value 0 1 2 5 10 30 60 120 300 600 1800 3600 seconds 2 After teh Sampling valu...

Page 23: ...nput Socket 3 12 Fig 1 The meter will permanent power ON when use the DC ADAPTER power supply The power Button function is disable 9 BATTERY REPLACEMENT 1 When the left corner of LCD display show it i...

Page 24: ...method During the power on use a pin to press the Reset Button 3 11 Fig 1 once a while will reset the circuit system 11 RS232 PC SERIAL INTERFACE The instrument has RS232 PC serial interface via a 3 5...

Page 25: ...it indicates the following status D15 Start Word D14 4 D13 When send the upper display data 1 When send the lower display data 2 D12 D11 Annunciator for Display 01 02 mW cm 2 A8 D12 A D11 8 D10 Polari...

Page 26: ...ra fast response naked bead thermocouple suitable for many general purpose application Thermocouple Measure Range 50 to 900 Probe 58 to 1650 Type K TP 02A Dimension 12cm tube 3 2mm Dia Thermocouple Me...

Page 27: ...PATENT The meter SD card structure already get patent or patent pending in following countries Germany Nr 20 2008 016 337 4 JAPAN 3151214 CHINA ZL 2008 2 0189918 5 ZL 2008 2 0189917 0 25 1111 UVC254S...

Reviews: