2-2 Electrical Specifications
A. Infrared Thermometer
Resolution/
1
蚓
- 20
蚓
to 650
蚓
ranges
1
蚌
-4
蚌
to 1202
蚌
Accuracy
< 400
蚓
3 % of reading or 3
蚓
( 5
蚌
),
which ever is greater.
400
蚓
3 % of reading.
*
Accuracy test under the measurement
range less than 300
蚓
( 572
蚌
).
* Meter operating Temp. within 23 5
蚓
& the emissivity value of measurement
target set to 0.95.
* Spec. tested with the 20 cm Dia. black
body, the measuring distance from the
probe sensing head is 30 cm.
* Spec. tested under the environment
RF Field Strength less than 3 V/M &
frequency less than the 30 MHz only.
Temp. Sensor
Thermocouple pie.
Emissivity
* By push button.
Setting
Setting range : 0.20 to 1.00.
* Factory preset emissivity value to 0.95,
which will cover 90% of a typical
application.
Measurement
6 to 12 micro meter.
Wave length
Region
Distance Factor
D/S : Approx. 7:1.
D - Distance, S - Spot.
4