Lutron Electronics SL-4011 Operation Manuals Download Page 6

4. MEASURING PROCEDURE

1. Slide the " A/C weighting selector " ( 3-4 ) to " A " or " C "
    position for sound level measuring.

Note :

a. The characteristic table of A, C weighting, please 

    ref. page 7.

b. The characteristic of A weighting is simulated as

    the " Human Ear Listening " response.

    Typically, if making the environmental sound level

     measurement, always select the A weighting typically.

c. The C weighting characteristic is near the " FLAT " 

    response. Typically, it is suitable for checking the

     noise of machinery ( Q.C. check ) & knowing the 

     real sound level of the tested equipment.

2. Determine proper measuring range by selecting the 
    " Range selector " ( 3-6 ) to minimize the tolerance of
    readout. When left corner of LCD show "   " 

    (Range upper / lower indicator, 3-9), it shows the dB 
    range selection is upper or lower setting. Slide range selector
    to other range for measuring.
3. According on various measuring sound source, select the 
    Time Weighting selector ( 3-5) to " Fast " or " Slow " position.
4. Hold the instrument in hand and point the microphone at
    measured noise source, the sound level will be displayed 
    on " dB " ( decibel) unit.
5. Max. hold - During the sound level measurement, if 
    need to store the maximum ( peak ) value on display, please
    slide the " Time weighting/ Max. hold selector " ( 3-5) to the
    " Max. hold " position.

    * When measure long-term stability under slowly varying 

       noise environment, please use the Max. hold function 

       to read the maximum values.

   * Push the " Reset Switch ( 3-11, Fig. 1 ) will reset the 

      Max. Hold value and execute the Max. Hold function 

      again.

4

Summary of Contents for SL-4011

Page 1: ...to the field of precision measurement Although this METER is a complex and delicate instrument its durable structure will allow many years of use if proper operating techniques are developed Please re...

Page 2: ...w Max hold selector 3 3 6 Range selector 3 3 7 Signal output terminal 3 3 8 Battery Compartment Cover 3 3 9 Range upper lower indicator 3 3 10 Calibration VR Accuracy adjust VR 3 3 11 Reset Switch for...

Page 3: ...d under load LCD display for low power consumption clear read out even in bright ambient light condition Used the durable long lasting components including a strong light weight ABS plastic housing ca...

Page 4: ...low response range are designed to meet IEC 61672 class 2 requirement Calibrator B K Bruel kjaer MULTIFUCTION ACOUSTIC CALIBRATOR 4226 Output Signal AC output AC 0 5 Vrms correspond ing to each range...

Page 5: ...witch Output type selector 3 4 A C weighting selector 3 5 Time Weighting Fast Slow Max hold selector 3 6 Range selector 3 7 Signal output terminal 3 8 Battery compartment Cover 3 9 Range upper lower i...

Page 6: ...e of readout When left corner of LCD show Range upper lower indicator 3 9 it shows the dB range selection is upper or lower setting Slide range selector to other range for measuring 3 According on var...

Page 7: ...rovided an 3 5 mm dia phone output 3 7 terminal for connection with analyzer level recorder tape recorder controller etc Slide Power switch Output type selector 3 3 to AC output or DC output according...

Page 8: ...ALIBRATOR model SC 941 or SC 942 Power on the Sound calibrator plug calibrator output into the Electric microphone 3 1 of the Sound Level Meter 2 Slide the Range selector 3 6 to 50 100 dB position 3 S...

Page 9: ...dB 2 5 dB 125 16 1 dB 0 2 dB 2 0 dB 250 8 6 dB 0 dB 1 9 dB 500 3 2 dB 0 dB 1 9 dB 1 K 0 dB 0 dB 1 4 dB 2 K 1 2 dB 0 2 dB 2 6 dB 4 K 1 dB 0 8 dB 3 6 dB 8 K 1 1 dB 3 dB 5 6 dB 10 TIME WEIGHTING FAST and...

Reviews: